pppppH CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

H CALIBRATION

For better accuracy, frequent calibration of the instrument is recommended.
The instrument should be recalibrated for pH:
a) Whenever the pH electrode or temperature probe is replaced
b) At least once a week
c)  After testing aggressive chemicals
d) When extreme accuracy is required

PREPARATION

Pour small quantities of pH 7.01 (

HI 7007

) and pH 4.01 (

HI 7004

) or pH

10.01 (

HI 7010

) buffer solutions into two clean beakers.

For accurate calibration use two beakers for each buffer solution, one for rinsing
the electrode tip, and one for calibration. In this way contamination of the
buffers is minimized.
For measurements in acidic samples, it is recommended to calibrate the meter
by using pH 7.01 (

HI 7007

) and pH 4.01 (

HI 7004

) buffers, while for

alkaline measurements use pH 7.01 (

HI 7007

) and pH 10.01 (

HI 7010

)

buffers.

PROCEDURE

• Connect the pH electrode and the temperature probe,

then switch the meter ON.

• Remove the electrode protective cap, rinse the electrode

tip with pH 7.01 solution, then immerse the pH
electrode and temperature probe into pH 7.01 buffer
solution;  stir gently and wait a few minutes for the
electrode to stabilize and reach thermal equilibrium.

Note:

The electrode should be submerged approximately 4
cm (1½") into the solution. The temperature probe
should be located close to the pH electrode.

• Press RANGE to display pH measurement.
• Press CAL to enter the calibration mode. The buffer value @25°C (77°F) and

the "pH" symbol will blink on the display.

The meter expects a pH 7.01 buffer.

20

HI 7082

3.5 M KCl electrolyte solution, 4 x 30 mL bottle, for double

junction electrodes

HI 8082

3.5 M KCl electrolyte solution, 4 x 30 mL FDA bottle, for double

junction electrodes

HI 8093

1 M KCl+AgCl electrolyte solution, 4 x 30 mL FDA bottle, for

double junction electrodes

ORP SOLUTIONS

HI 7091M

Reducing pretreatment solution, 230 mL bottle

HI 7091L

Reducing pretreatment solution, 500 mL bottle

HI 7092M

Oxidizing pretreatment solution, 230 mL bottle

HI 7092L

Oxidizing pretreatment solution, 500 mL bottle

HI 7020M

Test solution @200-275 mV, 230 mL bottle

HI 7020L

Test solution @200-275 mV, 500 mL bottle

HI 7021M

Test solution @240 mV, 230 mL bottle

HI 7021L

Test solution @240 mV, 500 mL bottle

HI 7022M

Test solution @470 mV, 230 mL bottle

HI 7022L

Test solution @470 mV, 500 mL bottle

pH ELECTRODES

HI 1043B / HI 1040S

Glass body, double junction, refillable, combination 

pH

 electrode.

Use: strong acid/alkali.

HI 1053B / HI 1050S

Glass body, triple ceramic, conic shape, refillable, combination 

pH

 electrode.

Use: emulsions.

B = BNC  CONNECTION PLUG + 1 m (3.3’) CABLE

S = SCREW-TYPE CONNECTOR

120 mm

4.7"

12 mm

0.5"

9.5mm DIA

0.37"

"S" VERSION

HI 1043

120 mm

4.7"

12 mm

0.5"

"S" VERSION

HI 1053

Summary of Contents for HI 8424NEW

Page 1: ...02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 273...

Page 2: ...e instru ment If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges in...

Page 3: ...itches from 0 1 mV to 1 mV when readings exceed 699 9 mV The user can set the following parameters by entering a setup mode through the keyboard enabling or disabling the auto off feature enabling or...

Page 4: ...S 5 Cheese FC 200B FC 200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1230B HI 1210S 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 1331B HI1310S 11 Foodindustrygeneralu...

Page 5: ...tempC pocket size thermometer 50 0 to 150 0 C HI 98502 ChecktempF pocket size thermometer 58 0 to 302 0 F HI 710015 Shockproof rubber boot blue HI 710016 Shockproof rubber boot orange HI 710022 Spare...

Page 6: ...isplay will show the pH value automatically compensated for temperature Notes In order to take accurate pH measurements make sure that the instrument has been calibrated before use see page 9 If measu...

Page 7: ...ase at 25 C ORP MEASUREMENTS Oxidation Reduction Potential ORP measurements provide a quantification of the oxidizing or reducing power of the sample tested Connect the ORP electrode optional to the B...

Page 8: ...HOLD again to return to normal mode If enabled keypresses are followed with an acoustic signal A lower note indicates that the key is not currently active To save battery life the meter is provided wi...

Page 9: ...ubmerged approximately 4 cm 1 into the solution The temperature probe should be located close to the pH electrode Press RANGE to display pH measurement Press CAL to enter the calibration mode The buff...

Page 10: ...19 pH CALIBRATION SOLUTIONS HI 70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI 7004M pH 4 01 buffer solution 230 mL bottle HI 7004L pH 4 01 buffer solution 500 mL bottle HI 8004L pH 4 01 buffer...

Page 11: ...er immerse it 4 cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution to be tested before...

Page 12: ...P to move to next step At this point it is possible to set the temperature unit by selecting Unit C or Unit F Press CAL SELECT to toggle the selection and CFM SETUP to exit the menu selection mode and...

Page 13: ...pH electrode Soak the electrode tip excessive drift in HI 7061 solution for 30 minutes Reading fluctuates Clogged dirty junction Soak the electrode tip up and down noise or low electrolyte level in wa...

Page 14: ...a good seal Replacement should take place in a non hazardous area using a 9V Ni MH rechargeable battery Note The meter can also be powered with a normal 9V alkaline battery NEVER USE THE HI 710040 CH...

Reviews: