Hanna Instruments HI 83740 Instruction Manual Download Page 4

4

The 

HI 83740 

is an auto-diagnostic portable microprocessor meter that benefits from Hanna’s

years of experience as a manufacturer of analytical instruments. It has an advanced optical system

based on a special tungsten lamp and a narrow band interference filter that allows most accurate

and repeatable readings. All instruments are factory calibrated.

The auto-diagnostic feature of this meter ensures always optimal measurement conditions to

ensure most precise readings. The light level is automatically adjusted each time a zero-measurement

is made, and the temperature of the lamp is controlled to avoid overheating.

GENERAL DESCRIPTION

Grapes accumulate normally only a small amount of copper by natural translocation from roots.

Unless exposed to significant airborne pollution or vineyard sprays, increased concentrations in wine

result from contamination during post-fermentation processing, like contact with non stainless steel

equipment and as impurities in fining agents and filter media.

The copper concentration in wine is normally low, less than 0.10 to 0.30 mg/L, because excess copper

is precipitated during fermentation due to adsorption onto the yeast cells. This adsorption and precipi-

tation can reduce the initial copper concentration with 40 to 89%. At higher concentration copper

plays an important role in catalysing oxidation reactions of wine phenols.

It is important to check the copper content both in must and in wine, because at levels above 9 mg/L

copper becomes a metabolic toxin that inhibits or delays alcoholic fermentation, and concentrations

exceeding 1 mg/L may be sensorial detected and should be avoided.

Other copper related problems can be manifested as formation of white haze (in white wines) and

later as a reddish-brown amorphous precipitate. This precipitated ‘casse’ develops only under the

strongly reducing conditions found in bottled wines. It has been found that this casse is a mixture of

copper compounds and proteins.

Factors favouring and inhibiting copper casse formulation in wine

Necessary conditions

Preventive

for copper casse formation

Measures

strong reducing conditions

copper levels at less than 0.3 mg/L

low iron concentrations

limit SO

2

 addition

high protein levels

cold-stabilize and bentonite fine to reduce proteins in white wine

light and heat

SIGNIFICANCE OF USE

17

Battery replacement must only take place in a non-

hazardous area.

The blinking “ ” will appear when the batteries power

gets low.

When batteries are completely discharged, “0% bAtt”

will appear and after two seconds the instrument is

switched off.

Remove the battery cover from the bottom of the

instrument and change the old batteries with 4 fresh

1.5V batteries, paying attention to the correct polarity.

Replace the cover.

BATTERIES REPLACEMENT

REAGENT SETS

HI 83740-20

Copper reagent set for wine (20 tests)

HI 83742-25

Color Reagent Set for wine (Wine Solvent-1)

OTHER ACCESSORIES

HI 740027P

1.5V AA batteries (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731321

Glass cuvets (4 pcs)

HI 731325W

Caps for cuvets (4 pcs)

HI 93703-50

Cuvets cleaning solution (230 mL)

HI 740231

20 mL glass cylinder with caps (2 pcs)

ACCESSORIES

Summary of Contents for HI 83740

Page 1: ...Instruction Manual HI 83740 COPPER ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83740R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at t...

Page 3: ...date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Inst...

Page 4: ...L copper becomes a metabolic toxin that inhibits or delays alcoholic fermentation and concentrations exceeding 1 mg L may be sensorial detected and should be avoided Other copper related problems can...

Page 5: ...nitions In a laboratory using a standard solution of 0 50 mg L copper and a representative lot of reagent an operatorobtainedwithasingleinstrumentastandard deviation of 0 05 mg L PRECISION AND ACCURAC...

Page 6: ...lts Block diagram optical layout 15 Press TIMER and the instrument will show the countdown or alternatively wait for 10 minutes leaving the vials capped and undisturbed During this period the color of...

Page 7: ...important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use the same cuvet for both It is necessary...

Page 8: ...n 7 READ key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE 20...

Page 9: ...uting procedure of the wine sample in case the L Lo Low Light message appears use the pipette to fill the glass vials with 5 mL of wine sample both vials then fill the vials up to the 15 mL mark with...

Page 10: ...lowed during testing to ensure best accuracy Use the supplied plastic pipette for adding the exact amount of wine sample 15 mL or reagents A or B up to the 20 mL mark to the graduated glass vials The...

Reviews: