Hanna Instruments HI 83740 Instruction Manual Download Page 10

10

ERROR  MESSAGES

The meter has lost its configuration. Contact your dealer or the nearest

Hanna Customer Service Center.

“Light high”: there is too much light to perform a measurement.

Please check the preparation of the zero cuvet.

“No Light”: the instrument cannot adjust the light level. Please check

that the sample does not contain any debris.

a) on zero reading:

A flashing value of the maximum concentration indicates an over range

condition. The concentration of the sample is beyond the programmed

range: dilute the sample and measure again.

The sample absorbs less light than the zero reference.

Check the procedure.

“Inverted”: the sample and the zero cuvet are inverted.

b) on sample reading:

Indicates that the batteries are dead and must be replaced. After this

message appears, the instrument is switched off. Change the batteries

and restart the meter.

“Light low”: there is not enough light to perform a measurement.

Dilute the sample. See “General tips for an accurate measurement“

(page 12).

11

The instructions listed below should be carefully followed during testing to ensure best accuracy.
• Use the supplied plastic pipette for adding the exact amount

of wine sample (15 mL) or reagents A or B (up to the 20

mL mark) to the graduated glass vials. The liquid in the

vial forms a convexity on the top; the bottom of this

convexity must be at the same level of the mark.

• To transfer the supernatant organic solvent from the vials to the cuvets use two different clean

1 mL pipettes, one for the zero and one for the sample.

a) squeeze the bulb of the pipette;

b) insert the plastic pipette into the supernatant organic solvent and release the bulb slowly,

paying attention not to transfer the wine sample too;

c) fill a cuvet with organic solvent by squeezing the bulb of the pipette.

Repeat step a), b) and c) until all the organic solvent is transferred. 

Ensure that at least 1/3

of the cuvet is filled with organic solvent, 

otherwise erroneous results will be obtained.

Notes:

- If some wine sample is transferred too, this does not interferes with the method.

- Try to avoid transferring suspended solids that might be present.

- If necessary, air bubbles can be removed tapping the vials gently on the table.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

Summary of Contents for HI 83740

Page 1: ...Instruction Manual HI 83740 COPPER ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83740R3 04 06 w w w h a n n a i n s t c o m...

Page 2: ...you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at t...

Page 3: ...date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Inst...

Page 4: ...L copper becomes a metabolic toxin that inhibits or delays alcoholic fermentation and concentrations exceeding 1 mg L may be sensorial detected and should be avoided Other copper related problems can...

Page 5: ...nitions In a laboratory using a standard solution of 0 50 mg L copper and a representative lot of reagent an operatorobtainedwithasingleinstrumentastandard deviation of 0 05 mg L PRECISION AND ACCURAC...

Page 6: ...lts Block diagram optical layout 15 Press TIMER and the instrument will show the countdown or alternatively wait for 10 minutes leaving the vials capped and undisturbed During this period the color of...

Page 7: ...important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use the same cuvet for both It is necessary...

Page 8: ...n 7 READ key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE 20...

Page 9: ...uting procedure of the wine sample in case the L Lo Low Light message appears use the pipette to fill the glass vials with 5 mL of wine sample both vials then fill the vials up to the 15 mL mark with...

Page 10: ...lowed during testing to ensure best accuracy Use the supplied plastic pipette for adding the exact amount of wine sample 15 mL or reagents A or B up to the 20 mL mark to the graduated glass vials The...

Reviews: