Oxidation-reduction potential (ORP) measurements provide the quantification of the oxidizing or reducing power of the
tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
DIRECT MEASUREMENT
To measure the mV of a sample using the Direct reading mode:
• Press
and then
to enter
mV Measure mode
(for the selected channel –
HI 4222
only).
• Select the Direct reading mode (see mV Setup for details, page 32).
• Submerge the tip of the ORP electrode (4 cm/1½”) into the
sample to be tested and allow a few seconds for the reading
to stabilize.
• The instrument will display the measured mV value on the LCD.
Note
: If the reading is out of range, “-----” will be displayed
on the LCD.
DIRECT / AUTOHOLD MEASUREMENT
To measure mV of a sample using the Direct/AutoHold reading mode:
• Press
and then
to select
mV Measure mode
(for the selected channel –
HI 4222
only).
• Select the Direct/AutoHold reading mode (see mV Setup for
details, page 32).
• Submerge the tip of the ORP electrode approximately 4
cm (1½”) into the sample to be tested and allow a few
seconds for the reading to stabilize.
• The measured mV value will be displayed on the LCD. If
pressing
, the “AutoHold” indicator will start blinking
on the LCD until the stability criterion is reached. The mV value
will be frozen on the LCD, along with “AutoHold” indicator.
• To return to normal
Measure mode press
.
Note
: If the reading is out of range, “-----” will be displayed
on the LCD.