To set the Electrode Type:
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the Electrode Type option.
• Press
and use
or
to select the desired
standard ISE or a custom one from the list.
For standard ISE:
• Press
to visualize the Ion constants and then press
at any time to exit Ion Constants view mode.
• Press
to confirm your selection and return to
ISE
Setup
options.
For custom ISE:
• Press
to edit the Ion constants for the selected
custom ISE. Use
or
to select the desired
Ion constant and press
to enter Ion constant edit
mode or
to cancel operation.
• For the Ion name the Text Editor menu will be displayed
on the LCD. Enter the desired information by accepting
the highlighted character which is added in the text bar,
using
. The
and
keys help the
user to select the desired character. It is also possible to
delete the last character by positioning the cursor on the
Backspace character and pressing
. Press
to return to the Ion Constants menu. If the Saving
Confirmation is enabled, press
to accept the
modified option,
to escape without saving or
to return to the editing mode. Otherwise, the
modified option is saved automatically.
• To set the appropriate Ion molar weight (in g/mol units)
use
or
and then press
to save
the current value or press
to cancel operation.