
2 5
Sample ID
This option allows the user to give to the measured samples an identification number/name. Two Sample ID
options are available: ID Increment Mode and Edit Sample ID.
ID Increment Mode
Two increment modes for the sample ID can be selected:
None
–
the sample ID will be fixed and it can be set alphanumerically.
Automatic
–
the sample ID will be increased with 1 from the set value, for each new log lot.
To set the ID Increment Mode:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Sample ID option.
• Press
and use
or
to highlight the ID
Increment Mode option.
• Press
and use
or
to highlight the
desired option.
• Press
to confirm your selection or press
to
cancel operation.
Edit Sample ID
This option allows the user to edit the sample ID.
Note
: If the ID Increment Mode is set to None, the sample ID can be set alphanumerically, otherwise it can be
set only numerically.
To edit the Sample ID:
• Press
while in
pH Measure
mode.
• Press
.
• Use
or
to select the Sample ID option.
• Press
and use
or
to highlight the Edit
Sample ID option.
• Press
to confirm your selection.
• If the selected increment mode is None, the Text Editor menu will be
displayed on the LCD, allowing you to enter the desired sample
number/name by accepting the highlighted character which is added
in the text bar, using
. The
and
keys help