background image

1

Instruction Manual

SENSOR

HANDLE

HI

4000-50

HI

41

14

POT

ASSIUM

COMBINA

TION

HI 4014

HI 4014

HI 4014

HI 4014

HI 4014

HI 4114

HI 4114

HI 4114

HI 4114

HI 4114

Potassium Ion

Selective Electrode

    Half-cell

    Combination

Summary of Contents for HI 4114

Page 1: ...ruction Manual SENSOR HANDLE HI 4000 50 HI 4114 POTASSIUM COMBINATION HI 4014 HI 4014 HI 4014 HI 4014 HI 4014 HI 4114 HI 4114 HI 4114 HI 4114 HI 4114 Potassium Ion Selective Electrode Half cell Combin...

Page 2: ...l n Charge on ion equivalents mol 1 A ion ion activity in sample T absolute temperature in K F Faraday constant 9 648 x 104 C equivalent HI 4014 P HI 4014 P HI 4014 P HI 4014 P HI 4014 Potassium half...

Page 3: ...ure otassium Measure otassium Measure otassium Measure ments ments ments ments ments 0 1 M potassium standard 500 mL HI 4014 01 HI 4014 01 HI 4014 01 HI 4014 01 HI 4014 01 Ionic Strength Adjuster 500...

Page 4: ...ore using During measurement always operate electrode with the fill hole open During normal use fill solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessivelo...

Page 5: ...shipping is in contact with inner membrane surface 4 Prepare reference electrode by filling outer electrolyte reservoir with fill solution HI 7076 5 Place sensor and reference electrodes into electro...

Page 6: ...tion Corrective Action Verifymodulehasbeenscrewedintosensorhandleor inner stem Verify seal has been removed from ceramic junction HI4114 orHI5315 referenceelectrode Verify electrodes are connected pro...

Page 7: ...ng after calibrating the meter with the standards HI 4014 00 ISE is added to both samples and standards to adjust the ionic strength of the solutions at a dose of 2 mL per 100 mL standard or sample Th...

Page 8: ...mbled Do not expose module to elevated temperatures Refrigeration of module will extend its life XV XV XV XV XV Conversion T Conversion T Conversion T Conversion T Conversion Tables ables ables ables...

Page 9: ...nthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplication t...

Reviews: