I. Introduction:
Introduction:
Introduction:
Introduction:
Introduction:
The Hanna HI 4009 and HI 4109 are ion selective elec-
trodes designed for the measurement of Cyanide ions in
aqueous solutions. The HI 4009 is a solid state half-cell
sensor that requires a separate reference. The HI 4109 is a
combination ion selective electrode.
II
IIII
IIII.
Specifications
Specifications
Specifications
Specifications
Specifications
Type:
Solid State electrode with
a Silver halide pellet.
Ion(s) measured:
Cyanide (CN
- - - - -
)
Measurement range: 10
-2
M to 1X 10
-6
M
260 to 0.26 ppm
Interfering ions:
Silver, Sulfide and
Mercury must be absent.
Ratio of interfering ion to
CN
-----
must be less than the
ratio indicated below:
1.0 for I
-----
Iodide
500 for Br
-----
Bromide
500 for Cl
-----
Chloride
Operating Temperature:
0-80
°
C
Operating pH:
11 to 14 pH
(recommended)
Dimensions:
12 mm (OD) X 120 mm
nominal insertion
(0.47
”
X 4.72
”
)
Connection:
BNC
WARNING:
HCN GAS IS TOXIC IF BREATHED, OR ABSORBED THROUGH
SKIN. USE OF A HOOD AND PROTECTIVE CLOTHING IS
STRONGLY ADVISED. USE ISA TO ADJUST pH ABOVE 11.