Hanna Instruments HI 4109 Instruction Manual Download Page 14

14

A cyanide electrode may be used as an indicator to follow
the progress and detect the endpoint of a complexation
titration of CN

-

 

with silver nitrate standard. Time must be

allowed for the alkali salt to redissolve and mV to settle.
The end point is reached when the formation of insoluble
silver cyanoargentate precipitates occurs.  The electrode can
be used in colored samples, or high or variable ionic strength
samples to increase accuracy of the determination. During
the titration the sensor follows the decrease in cyanide
concentration while small additions of silver nitate titrant
are added. The silver reacts with the cyanide ions forming
a precipitate.
Ag

+

 + 2CN

-

 

<—>

 [Ag(CN)

2

]

-   

(soluble)

[Ag(CN)

2

]

<—>

 Ag

2AgCN 

(insoluble)

At the stoichiometric end point, a large change in mV
occurs. This technique may be useful in obtaining the actual
value of cyanide standards. Measurements may be
automated by use of the Hanna Titrator HI 901 or titrated
manually.

XIII.

XIII.

XIII.

XIII.

XIII. pH

 pH

 pH

 pH

 pH

The HI 4109 and HI 4009 electrodes  may be used in
solutions with pH values between 11 and 13.  Samples
that fall beyond this range should be adjusted.

XIV. S

XIV. S

XIV. S

XIV. S

XIV. Storage and Care of the HI 4009 and

torage and Care of the HI 4009 and

torage and Care of the HI 4009 and

torage and Care of the HI 4009 and

torage and Care of the HI 4009 and

     

     

     

     

      HI 4109 sensors

 HI 4109 sensors

 HI 4109 sensors

 HI 4109 sensors

 HI 4109 sensors

The HI 4009 sensor can be stored in very dilute standards
(<10

-3

M) for short periods of time and should be stored

dry with the protective cap on when not in use.
The model HI 4109 combination electrode can be left in
dilute standards (<10

-3

 M) for short time periods.

For long term storage, the electrode should be drained and
washed of salts with distilled or deionized water. Unscrew
the upper cap and move outer sleeve up cable.  Wrap the
ceramic junction on the inner stem with Parafilm

®

 or other

sealing wrap. Place the protective cap provided over the

Titration

Summary of Contents for HI 4109

Page 1: ...1 HI 4009 HI 4009 HI 4009 HI 4009 HI 4009 HI 4109 HI 4109 HI 4109 HI 4109 HI 4109 Cyanide Ion Selective Electrode Instruction Manual Half cell Combination CYANIDE HI 4009 HI 4109 CYANIDE COMBINATION...

Page 2: ...uires a separate reference The HI 4109 is a combination ion selective electrode II II II II II Specifications Specifications Specifications Specifications Specifications Type Solid State electrode wit...

Page 3: ...te itselectrolyticcircuit TheHI4109incorporates areference electrode The silver halide pellet is practically insoluble in thetestsolutionsbeingmeasuredandproducesapotential change due to changes in th...

Page 4: ...HI 4109 Design elements of the HI 4009 and HI 4109 Design elements of the HI 4009 and HI 4109 electrodes electrodes electrodes electrodes electrodes Upper Cap Spring O Ring Ceramic Junction on Inner S...

Page 5: ...rements for Cyanide Measurements ISA 500 mL HI4001 00 User supplied standards Note Thesesolutions shouldonlybeusedformakingmore dilute standards The sensing surface will be damaged if submersed in thi...

Page 6: ...by the corresponding dilution factor For high ionic strength samples use standard addi tion or titration methods Calibration standards and sample solutions should be at same temperature and at elevat...

Page 7: ...nce fill solution should be added daily toelectrolytereservoirbeforeelectrodeuse During measurement always operate electrode with the fill hole open During normal use fill solution will slowly drain o...

Page 8: ...lm seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Page 9: ...nd verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just belo...

Page 10: ...Corrective action Corrective action Corrective action Corrective action Corrective action Verify protective cap has been removed HI 4009 Verify plastic film has been removed from inner stem HI4109 Ve...

Page 11: ...brated with two or more freshly made standards that are in the linear measurement range of the unknowns One mL of ISA HI 4001 00 is added to each 100 mL volume of standard or sample More calibration s...

Page 12: ...ix before taking measurements 3 Follow sectionVII GeneralGuidelinestooptimizetest set up 4 Duringcalibrationitisbesttostartwithlowerconcen tration samples first Wait for a stable measurement before re...

Page 13: ...pH ISE mV meter which simplifies the method greatly The method is recommended for samples with higher ionic strengths Example Cyanide ion determination in samples with con centrations less than 5 X 1...

Page 14: ...the Hanna Titrator HI 901 or titrated manually XIII XIII XIII XIII XIII pH pH pH pH pH The HI 4109 and HI 4009 electrodes may be used in solutions with pH values between 11 and 13 Samples that fall b...

Page 15: ...Disposal of Cyanides Disposal of Cyanides Sodium Cyanide is a deadly poison and extreme care must be taken during use and in disposal after use Test solutions and standards may be neutralized before d...

Page 16: ...r 6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplica...

Reviews: