XIII. 

XIII. 

XIII. 

XIII. 

XIII. Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

Direct Calibration and Measurement

The direct method can be used in the linear working re-
gions of the sensor. (See figure for typical sensor response).
A direct reading ISE meter (HI 4222 or equivalent) deter-
mines concentration of the unknown by a direct reading
after calibrating the meter with the standards. The meter is
calibrated with two or more freshly made standards that
are in the measurement range of the unknowns. HI 4001-
00 ISA is added just before measurement of the  standard
or sample. Covering the vessel to prevent gas loss is ad-
vised.
A pH/mV meter in mV mode and semi-log graph paper
may also be used. Two or freshly prepared standards that
are in the measurement range of the unknowns (with ISA
added), are measured in mV mode on the meter.  These
standards are plotted on semilog graph paper and their
points are connected to form a straight-line curve.  When
samples are measured, their mV values are converted to
concentration by following the mV to the concentration axis
on the semi-log plot.
Method 4500-NH

3

 D. from Standard Methods for the Ex-

amination of Water and Wastewater is a direct measure-
ment method for water samples.
For both direct reading and mV convertion, ISA is added
prior to measurement and the vessel should be covered to
prevent gas loss.
In the lower concentration ranges the electrode calibration
becomes less linear, many more calibration points are
needed, and calibration will need to be repeated more
frequently.  Known addition method may also be used in
these regions provided the actual slope of the sensor has
been determined.
Direct Measurement Procedure
1)      Follow section IX to prepare sensor.
2)      Follow section VI to prepare standards and solutions.

Summary of Contents for HI 4101

Page 1: ...1 HI 4101 HI 4101 HI 4101 HI 4101 HI 4101 Ammonia Ion Selective Electrode Instruction Manual A M M O N IA HI 4101...

Page 2: ...2...

Page 3: ...monium ions are also measured by conver sion to ammonia gas upon ISA addition II II II II II Specifications Specifications Specifications Specifications Specifications Type NH3 gas sensing electrode w...

Page 4: ...gas in the sample and thin film are equal The change in pH is proportional to the concentra tion of dissolved gas in the sample solution The Nernst expression for an ammonia sensor is expressed in th...

Page 5: ...ethemembrane membranecap outerprobe body with antirotation key and the pH reference assembly which includes the outer electrode cap spring inner cap and pH reference electrode assembly Outer probe bod...

Page 6: ...A HI4001 00 Hanna 0 1 M standard HI4001 01 Hanna 100 ppm N standard HI4001 02 Hanna 1000 ppm N standard HI4001 03 Please Note These calibration standards are ppm as NH3 N See Section XVII for addition...

Page 7: ...atingor wetting thePTFEmembranewill effect the response Inspect sensor before using Re place PTFE membrane if damage is evident Rinse electrode with distilled or deionized water be tweensamplesanddabd...

Page 8: ...ionized water These pH solutions contain chloride ions and pH buffers that are used to verify the inner electrode pH internal is operational See Section XVII for replacement accessories and maintenanc...

Page 9: ...lution carry over Do not rub the glass Take a measurement in the second buffer and record mV Pay attention to minus sign if present Calculate the difference in mV between the two solutions Example of...

Page 10: ...le gently stretching membrane over opening and cap turing opposite membrane corner against threads Smoothexcessmembranematerialaroundthethreads 4 Screw outer membrane cap onto body thus capturing the...

Page 11: ...de spring and elec trode cap down cable and screw cap on outer body until fully engaged Do not invert electrode Do not overtighten 8 Install assembled electrode in gas sensor test vessel or in electro...

Page 12: ...has stabilized This value should be less than the previous noted more negative DeterminethedifferencebetweenthetwomVvalues An acceptable value for this slope is 54 4 mV at ambient temperatures between...

Page 13: ...bound nitrogen such as oil sludge waste or samples which may contain surfactants digest sample first using a total Kjeldahl nitrogen TKN procedure This involves oxidationwithhotsulfuricacidwhich conv...

Page 14: ...in the measurement range of the unknowns with ISA added are measured in mV mode on the meter These standards are plotted on semilog graph paper and their points are connected to form a straight line c...

Page 15: ...ed gas by covering and using promptly 3 FollowsectionVII GeneralGuidelinestooptimizetest set up 4 Duringcalibrationitisbesttostartwithlowerconcen trationsamplesfirst Waitforastablereadingbefore readin...

Page 16: ...erature of measurement should be determined and used if possible This will improve ac curacy Known addition is Method 4500 NH3 E from StandardMethodsfortheExaminationofWaterandWaste water 1 The volume...

Page 17: ...with an electrode 1 mL of ISA is added to the sample and the color change is noted The sample is covered and permitted to mix The mV is then recorded when the sensor has stabilized 2 5 mL VSTANDARD o...

Page 18: ...mpletely and rinse off the internal pH reference assembly the outer body and the membrane cap Discard the white PTFE membrane Note keep black membrane cap Cover the glass tip with the protective shipp...

Page 19: ...ppm N standard 500 mL HI4001 03 Hanna 1000 ppm N standard 500mL HI4001 30 Nitrate test kit Bulk pkg Please Note These calibration standards are ppm as NH3 N For Maintenance For Maintenance For Mainten...

Page 20: ...m date of purchase when used for their intended purpose and maintained according to instructions If they fail to work whenfirstusedcontactyourdealerimmediately Damagedue to accidents misuse misapplica...

Reviews: