For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

If the filling solution (electrolyte) is more than 2½ cm (1") below the
fill hole, add 

HI 7082

 or 

HI 8082

 3.5M KCl Electrolyte Solution for

double junction or 

HI 7071

 or 

HI 8071

 3.5M KCl+AgCl Electrolyte

Solution for single junction electrodes.
Unscrew the fill hole screw during measurements.

For AmpHel

For AmpHel

For AmpHel

For AmpHel

For AmpHel

®

®

®

®

®

 electrodes:

 electrodes:

 electrodes:

 electrodes:

 electrodes:

If the electrode does not respond to pH changes, the battery is dead
and the electrode should be replaced.

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

Rinse the electrode tip with distilled water. Immerse the tip (bottom
4 cm /1½”) into the sample and stir gently for a few seconds.
For a faster response and to avoid cross-contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.

STORAGE 

STORAGE 

STORAGE 

STORAGE 

STORAGE  PROCEDURE

PROCEDURE

PROCEDURE

PROCEDURE

PROCEDURE

To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of 

HI 70300

or 

HI 80300

 Storage Solution or, in its absence, Fill Solution (

HI

7071

 or 

HI 8071

 for single junction and 

HI 7082

 or 

HI 8082

 for

double junction electrodes). Follow the Preparation Procedure on page
16 before taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.

PERIODIC MAINTENANCE

PERIODIC MAINTENANCE

PERIODIC MAINTENANCE

PERIODIC MAINTENANCE

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable must be intact and
well connected. No cracks should be seen on the electrode stem or
bulb. Connectors must be perfectly clean and dry. If any scratches or
cracks are present, replace the electrode. Rinse off any salt deposits
with water.

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

For refillable electrodes:

Refill the reference chamber with fresh electrolyte (

HI 7071

 or 

HI 8071

for single junction or 

HI 7082

 or 

HI 8082

 for double junction

electrodes). Allow the electrode to stand upright for 1 hour.
Follow the Storage Procedure above.

ELECTRODE CONDITIONING &

ELECTRODE CONDITIONING &

ELECTRODE CONDITIONING &

ELECTRODE CONDITIONING &

ELECTRODE CONDITIONING &

MAINTENANCE

MAINTENANCE

MAINTENANCE

MAINTENANCE

MAINTENANCE

PREPARATION PROCEDURE

PREPARATION PROCEDURE

PREPARATION PROCEDURE

PREPARATION PROCEDURE

PREPARATION PROCEDURE

Remove the protective cap of the pH electrode.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes. They will disappear when rinsed with water.
During transport, tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by “shaking down” the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in 

HI 70300

 or

HI 80300

     Storage Solution for at least one hour.

Summary of Contents for HI 2211

Page 1: ...1 Instruction Manual Microprocessor based pH mV C Bench Meters HI 2210 HI 2211 www hannainst com...

Page 2: ...pping any instrument make sure it is properly packed for complete protection WARRANTY 3 PRELIMINARY EXAMINATION 4 GENERAL DESCRIPTION 4 FUNCTIONAL DESCRIPTION 5 SPECIFICATIONS 6 OPERATIONAL GUIDE 7 pH...

Page 3: ...mV and temperature simultaneously together with graphic symbols The calibration process is guided step by step through clear indications on the LCD A stability indicator makes the calibration procedur...

Page 4: ...probe approximately 4 cm 1 into the sample to be tested and stir gently Allow time for the electrode to stabilize The pH is displayed on the primary LCD and the temperature on the secondary LCD If the...

Page 5: ...temperature probe into the sample as close as possible to the electrode and wait for a few seconds If the temperature of the sample is known manual temperature compensation can be performed by disconn...

Page 6: ...in this situation CAL 2 BUF pH BUFFER pH p p p p pH CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION Calibrate the instrument often especially if high accuracy is required The inst...

Page 7: ...down the CAL MEM keys then power on the instrument The CAL message will appear and the secondary LCD will show 0 0 C Immerse the temperature probe in the vessel with ice and water as near as possible...

Page 8: ...9 3 0 4 4 8 6 9 9 6 1 1 9 2 9 9 0 4 4 0 1 4 0 4 4 8 6 8 9 6 7 0 9 8 8 9 5 4 3 1 1 5 0 4 3 8 6 8 9 6 4 0 9 5 8 9 0 5 2 2 1 6 0 4 3 8 6 8 9 6 1 0 9 2 8 9 5 5 1 3 1 8 0 4 4 8 6 8 9 6 9 9 8 9 7 9 0 6 0 4...

Page 9: ...e Preparation Procedure on page 16 before taking measurements Note NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER PERIODIC MAINTENANCE PERIODIC MAINTENANCE PERIODIC MAINTENANCE PERIODIC MAI...

Page 10: ...s e g n a r e t y l o r t c e l e e h t k c e h C e l a r e n e g e h t d n a l e v e l e d o r t c e l e e h t f o e t a t s s w o h s y a l p s i d e h T d n a V m e h t n i e g n a r f o t u O e l...

Page 11: ...n FDA approved bottle 4x50 mL for single junction electrodes HI 8072 1M KNO3 Electrolyte in FDA approved bottle 4x50 mL HI 8082 3 5M KCl Electrolyte in FDA approved bottle 4x50 mL Typical Electrode Li...

Page 12: ...IA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution 460 mL bottle HI 7092L Oxidizing Pretreatment Solution 460 mL bottle pH ELECTRODES All electrodes part numb...

Page 13: ...I 3230 S VERSION 120 mm 4 7 12 mm 0 5 HI 4430 S VERSION FC 100B Plastic body PVDF double junction refillable combination pH electrode Use general purpose for food industry FC 200B Plastic body PVDF op...

Page 14: ...To avoid damage or burns do not perform any measurement in microwave ovens EXTENSION CABLE FOR SCREW TYPE ELECTRODES SCREW TO BNC ADAPTER HI 7855 1 Extension cable 1 m 3 3 long HI 7855 3 Extension cab...

Page 15: ...2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2...

Reviews: