27
Setup mode allows viewing and modifying the following parameters:
• Calibration Alarm Time Out
• One-point calibration behaviour
• Custom Buffer 1
• Custom Buffer 2
• Alarm High Limit
• Alarm Low Limit
• AutoLOG Start Condition
• AutoLOG End Condition
• AutoLOG Interval
• Temperature Unit
• Current Date (yyyy.mm.dd)
• Current Time (hh:mm)
• Printing Language (
HI 122
only)
• Printer Status (
HI 122
only)
• Beep Status
• Baud Rate (serial communication)
• Command prefix (serial communication)
• Instrument ID
To enter SETUP mode, press
2nd
then
SET
key, or simply
SET
from the left
keyboard (
HI 120
only), while the instrument is in measurement mode.
Select the desired setup parameter using the
ARROW
keys.
Press
CAL
if you want to change the item value. The selected item (e.g.
hour, in setting up the correct time) and “CFM” tag will start blinking.
Press the
ARROW
keys to change the displayed value or simply use the
numerical keys for all numerical parameters.
Note:
To use the numerical keys, press
NUM
from the left keyboard. The “2nd”
tag will blink. Set the desired number digit by digit. The new introduced
digit is always the last one. All the previous digits will shift to left.
To delete digits press
CLR
. Last introduced digit will be deleted and
the number will shift to right.
S E T U P