background image

INSTRUCTION MANUAL

Hanna Instruments Inc., 584 Park East Drive, Woonsocket, RI 02895 USA 

www.hannainst.com

HI9810452

Wireless pH Tester for Meat

with built-in specialized electrode

Foodcare

Summary of Contents for HALO2 HI9810452

Page 1: ...INSTRUCTION MANUAL Hanna Instruments Inc 584 Park East Drive Woonsocket RI 02895 USA www hannainst com HI9810452 Wireless pH Tester for Meat with built in specialized electrode Foodcare...

Page 2: ...8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 8 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8...

Page 3: ...05 pH mV Range pH mV conversion Resolution 0 1 or 1 mV Temperature Range 0 0 to 60 0 C 32 0 to 140 0 F Resolution 0 1 C 0 1 F Accuracy 0 5 C 0 9 F Calibration Up to three points or four points Automa...

Page 4: ...e g solvents sodium hypochlorite ultraviolet light and fungal growth the PVDF body has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and sem...

Page 5: ...nitialization screen displays all LCD segments followed by the battery percentage The tester enters measurement mode Before the tester turns off OFF PWR is displayed briefly Press and hold the ON OFF...

Page 6: ...ff 8 AOFF 60 AOFF AOFF Calibration Points Option 2P or 3P Press ON OFF button to select between up to two or up to three point calibration 2P CAL 3P CAL Buffer Set Option 7 01 pH Hanna or 6 86 pH NIST...

Page 7: ...t of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the applicatio...

Page 8: ...return to measurement mode press the ON OFF button CAL SAVE is displayed briefly With 2P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibrat...

Page 9: ...and improve the pH measurement Note Never immerse the tester over the maximum immersion level Refilling the Electrode To remove the electrode sleeve carefully rotate and slide it off A Set aside Note...

Page 10: ...r solution 20 mL sachet 25 pcs HI70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI70009P pH 9 18 buffer solution 20 mL sachet 25 pcs HI70010P pH 10 01 buffer solution 20 mL sachet 25 pcs HI77400P...

Page 11: ...and maintained according to instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance is not covered If...

Page 12: ...C 12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must...