7 Bluetooth
7. BLUETOOTH
With “
PAIr
BLU
” or “
On
BLU
” selected in Setup, the Bluetooth icon (
&
) will blink for up to 45 seconds,
indicating the tester is in discoverable mode. Once connected, the icon will stop blinking. If not connected,
the icon is not displayed.
• Select “
On
BLU
”, to enable Bluetooth without bonding.
• Select “
PAIr
BLU
”, to enable Bluetooth with bonding. A 6-digit bonding pin is displayed the first
time the tester and smart device are paired. Once the devices are paired, the bonding pin is not required
when reconnected.
• Select “
OFF
BLU
”, to disable Bluetooth.
• Select “
dEL
PAIR
”, to delete all paired devices. If PAIr BLU is enabled, a bonding pin will need to
be re-entered.
8. HANNA LAB APPLICATION
• The Hanna Lab App is available on the App Store
®
and on Google Play.*
• Consult the Help section of the application for information on calibration, measurement, data logging,
and sharing.
• When the tester is in discoverable mode, it will appear in the list of “Available Devices”.
• Within the application, tap “Connect” to pair the tester with the device. All readings are transmitted
directly to the application.
Tag a Measurement
Once connected to the application, the ON/OFF button can be pressed to tag the current reading.
• Press the ON/OFF button from measurement mode. The display shows “
SEt
TAG
”, followed by “
-
TAG
”.
• The reading on the application will flash green and the note icon ( ) will be displayed. Tap the note
icon ( ) to add an annotation.
&
10.12
25.6
&
-
TAG
9. CALIBRATION
Preparation & Guidelines
1. Remove the electrode from the plastic storage cap. Save the cap for electrode storage.
2. Rinse off any storage solution or salts that may be on the body.
3. Verify there is solution inside the pH bulb by shaking the electrode down to restore continuity as the
solution may have moved up the stem during shipping.
4. For best results, use a rinse beaker and a separate calibration beaker for each buffer. Discard rinsing
buffers after use.
Note
: Remove the blade prior to calibration and clean the electrode.