Hanna Instruments HALO2 HI9810352 Instruction Manual Download Page 7

7

BLUET
OO

Th

Bluetooth Pairing

Option:

 dEL PAIr

Press ON/OFF button to delete saved paired 

device.

&

dEL

PAIR

Date & Time

Option:

 SEt TIME

Press ON/OFF button to set the date and time.

Option:

 YEAR, MO, DAY, HOUR, and MIN

2021

YEAR

 

3

MO

 

15

DAY

Use the Setup button to select the option and press the ON/OFF button to change the selected option.

7.  BLUETOOTH

With “

PAIr

 

BLU

” or “

On

 

BLU

” selected in Setup, the Bluetooth icon (

&

) will blink for up to 45 seconds, indicating the 

tester is in discoverable mode. Once connected, the icon will stop blinking. If not connected, the icon is not displayed.

•  Select “

On 

BLU

”, to enable Bluetooth without bonding.

•  Select “

PAIr 

BLU

”, to enable Bluetooth with bonding. A 6-digit bonding pin is displayed the first time the tester and 

smart device are paired. Once the devices are paired, the bonding pin is not required when reconnected. 

•  Select “

OFF 

BLU

”, to disable Bluetooth.

•  Select “

dEL

 

PAIR

”, to delete all paired devices. If PAIr BLU is enabled, a bonding pin will need to be re-entered.

8.  HANNA LAB APPLICATION

•  The Hanna Lab App is available on the App Store

®

 and on Google Play.

1

 

•  Consult the Help section of the application for information on calibration, measurement, data logging, and sharing.

•  When the tester is in discoverable mode, it will appear in the list of “Available Devices”.

•  Within the application, tap “Connect” to pair the tester with the device. All readings are transmitted directly to the application.

Tag a Measurement

Once connected to the application, the ON/OFF button can be pressed to tag the current reading.

•  Press the ON/OFF button from measurement mode. The display shows “

SEt

 

TAG

”, followed by “

-

 

TAG

”.

•  The reading on the application will flash green and the note icon ( ) will be displayed. Tap the note icon ( ) to add 

an annotation.

&

10.12

25.6

   

&

-

TAG

9.  CALIBRATION

Preparation & Guidelines

1.  Unscrew the top collar to vent the cap before removing the probe from the plastic storage tube. Save the cap for probe 

storage.

2.  Rinse off any storage solution or salts that may be on the body.

3.  Verify there is solution inside the pH bulb by shaking the probe down to restore continuity as the solution may have 

moved up the stem during shipping.

4.  For best results, use a rinse beaker and a separate calibration beaker for each buffer. Discard rinsing buffers after use.

Procedure

For most applications it is recommended to start with pH 7.01 buffer (or pH 6.86).

To restore factory defaults, press and hold the ON/OFF button. “

CAL

 

CLR

” is displayed.

Note:

 It is recommended to calibrate the electrode with buffers at the temperature it will be used at.

App Store is a trademark of Apple, Inc.

Google Play and the Google Play logo are trademarks of Google LLC.

Summary of Contents for HALO2 HI9810352

Page 1: ...INSTRUCTION MANUAL HI9810352 Wireless pH Tester for Sushi with built in specialized electrode Foodcare ...

Page 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Page 3: ...Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Sushi Rice Measurement 8 Preparation 8 Direct Measurement 9 Measurement of Rice Slurries 9 11 Care Maintenance 9 Storage 9 12 Warning Error Mess...

Page 4: ...ion pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Titanium Glass Low Temperature LT Junction Open Reference ce...

Page 5: ... titanium works as an electronic shield protecting against interferences from electrical noise or humidity The flat glass tip provides optimal surface contact for sushi rice pH measurements and for surfaces that cannot be penetrated The open junction reference resists clogging due to the hard gel surface viscolene that is used for the reference cell When the junction becomes coated with starch fro...

Page 6: ...ed battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure menu item ...

Page 7: ... in discoverable mode it will appear in the list of Available Devices Withintheapplication tap Connect topairthetesterwiththedevice Allreadingsaretransmitteddirectlytotheapplication Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG followed by TAG The reading on the...

Page 8: ...buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 Place the tip of the electrode in the third buffer rinse beaker then in the third calibration buffer Wait until the measurement is stored and the stability indicator disappears 8 After the third point has been stored CAL SAVE is displayed briefly and the tester wi...

Page 9: ...on level Fresh buffers should be used for each calibration Once the sachets are opened the buffer value can change over time If the electrode is slow or sluggish soak it in cleaning solution for 20 minutes Rinse with water and hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating If measurements are taken successively rinse the probe thoroughly in distilled or de...

Page 10: ...y suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810352 is warranted for a period of one year against defects in workmanship and materials when used for its intended purp...

Page 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de li...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810352 Printed in ROMANIA ...

Reviews: