Hanna Instruments HALO2 HI9810352 Instruction Manual Download Page 6

6

GENE

rAL

 O

PE

rA

Ti

ONS

5.  GENERAL OPERATIONS

Turning the Tester On & Off

•  Turn the tester face down. Rotate the battery cover on the back of the tester counterclockwise and set it aside. Remove 

the battery insulation film. 

•  Press the ON/OFF button to turn the tester on or off. Initialization screen displays all LCD segments, followed by the 

battery percentage. The tester enters measurement mode. Before the tester turns off, “

OFF

 

PWR

” is displayed briefly.

•  Press and hold the ON/OFF button to turn the tester off when connected to Bluetooth.

Battery Replacement

1.  Turn off the tester. Turn the tester face down and rotate the battery cover counterclockwise. 

2.  Set the battery cover aside. Press the metallic pin to push old battery out.

3.  Place the new battery with positive (+) sign facing out.

4.  Reset the date and time in setup, or connect to the Hanna Lab App to update it automatically.

5.  Align the mark on the cover with the open lock icon ( ) on the case. Rotate the cover 

clockwise until the mark on the cover aligns with the closed lock icon ( ).

Note:

 Only use specified battery type. Dispose of old battery in accordance with local regulations.

6.  SETUP

The Setup button is located inside the battery compartment. After Setup configuration, replace the cover.

Setup Menu Navigation

•  Press the Setup button to enter Setup mode and navigate menu items.

•  To exit Setup mode, press the Setup button after “

SEt

 

TIME

” option is displayed.

•  Press the ON/OFF button to configure menu item options.

Temperature Unit 

Option:

 °C or °F

Press ON/OFF button to select desired 

temperature unit.

°C

T E M P

 

°F

TEMP

Auto-Off Interval

Option:

 8, 60 min., or “

----

” (disabled)

Press ON/OFF button to select desired interval. 

To save battery life, after the selected auto-off

8

AOFF

 

60

AOFF

 

----

AOFF

interval has elapsed, the tester will automatically turn off. 

Calibration Points

Option:

 2P or 3P

Press ON/OFF button to select between up to 

two- or up to three-point calibration.

2P

CAL

 

3P

CAL

Buffer Set

Option:

 7.01 pH (Hanna) or 6.86 pH (NIST)

Press ON/OFF button to select the calibration 

buffer set (Hanna or NIST).

7.01

BUFF

 

6.86

BUFF

pH Resolution

Option:

 0.01 pH or 0.1 pH

Press the ON/OFF button to select resolution.

0.01

RES

 

0.1

RES

 

Bluetooth Wireless Mode

Option:

 On, PAIr, or OFF

Press ON/OFF button to select Bluetooth 

option at start-up.

&

On

BLU

 

&

PAIr

BLU

 

&

OFF

BLU

Metallic pin

Summary of Contents for HALO2 HI9810352

Page 1: ...INSTRUCTION MANUAL HI9810352 Wireless pH Tester for Sushi with built in specialized electrode Foodcare ...

Page 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Page 3: ...Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Sushi Rice Measurement 8 Preparation 8 Direct Measurement 9 Measurement of Rice Slurries 9 11 Care Maintenance 9 Storage 9 12 Warning Error Mess...

Page 4: ...ion pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Titanium Glass Low Temperature LT Junction Open Reference ce...

Page 5: ... titanium works as an electronic shield protecting against interferences from electrical noise or humidity The flat glass tip provides optimal surface contact for sushi rice pH measurements and for surfaces that cannot be penetrated The open junction reference resists clogging due to the hard gel surface viscolene that is used for the reference cell When the junction becomes coated with starch fro...

Page 6: ...ed battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure menu item ...

Page 7: ... in discoverable mode it will appear in the list of Available Devices Withintheapplication tap Connect topairthetesterwiththedevice Allreadingsaretransmitteddirectlytotheapplication Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG followed by TAG The reading on the...

Page 8: ...buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 Place the tip of the electrode in the third buffer rinse beaker then in the third calibration buffer Wait until the measurement is stored and the stability indicator disappears 8 After the third point has been stored CAL SAVE is displayed briefly and the tester wi...

Page 9: ...on level Fresh buffers should be used for each calibration Once the sachets are opened the buffer value can change over time If the electrode is slow or sluggish soak it in cleaning solution for 20 minutes Rinse with water and hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating If measurements are taken successively rinse the probe thoroughly in distilled or de...

Page 10: ...y suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810352 is warranted for a period of one year against defects in workmanship and materials when used for its intended purp...

Page 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de li...

Page 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810352 Printed in ROMANIA ...

Reviews: