Unpack the instrument and place it on a flat surface. Do not place the instrument in dirrect sunlight.
Insert the 12 Vdc power plug into its input on the back of the instrument.
Switch on the instrument. On the LCD, the Hanna Logo will appear for a short time, followed by the
main screen for turbidity measurements.
The instrument loads the selected language. If no language can be loaded the instrument will work
in the
“safe mode”
. In
“safe mode”
all the messages are displayed in English and
tutorial and help information are not available.
HEATING THE SAMPLE
This procedure can be very effective however it should be handled with care because it can alter the
turbidity of the sample. When heating a sample, the volatile components from the sample can
vaporize, the suspended components can dissolve or the sample characteristics can change.
The heating procedure should be used with extreme care.
Use a warm water bath and immerse the cuvette filled with sample into the bath. Heat the sample
only until the visible bubbles are removed.
Note
: Always cool the heated sample to the original temperature before measurement.
The heating procedure can be used in combination with vacuum or ultrasonic waves application for
a more effective air bubble removal.