10. CALIBRATION
10.1. EC CALIBRATION
The
HI981413
provides a digital calibration at the push of a button. Calibrate the probe frequently
for improved accuracy. Also:
• before in-line or flow cell installation
• whenever the probe is replaced
• after periodic maintenance
Always use fresh calibration solutions and perform electrode maintenance prior to calibration (see
ELECTRODE CARE & MAINTENANCE section).
Preparation
Pour standard solution into clean beakers sufficient to cover the sensing portion of the probe (1
1/2
").
If possible, use plastic beakers to minimize any EMC interferences. For accurate calibration and to
minimize cross-contamination, use two beakers, one for rinsing the probe and one for calibration.
1 1/2 “
Procedure
• Raise and lower the probe in the standard to ensure the entire cell area is filled with standard.
• Shake any bubbles off the two electrodes (prongs),
• Center probe in beaker away from beaker walls.
Automatic One-Point Calibration
One-point calibration can be performed using one of the standard solutions: 1.41 or 5.00 mS/cm.
• Press CAL key to enter calibration mode. “mS/cm 1.41 USE” message is displayed as the
calibration point. The value will change to 5.00 if that standard is used.
• The controller automatically recognizes the standard value.