INSTRUCTION MANUAL

HI9810302

Wireless pH Tester 

for Soil

with built-in specialized electrode

All manuals and user guides at all-guides.com

Summary of Contents for GroLine HALO2 HI9810302

Page 1: ...INSTRUCTION MANUAL HI9810302 Wireless pH Tester for Soil with built in specialized electrode All manuals and user guides at all guides com...

Page 2: ...oduct Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise ide...

Page 3: ...s Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibrati...

Page 4: ...esolution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard...

Page 5: ...t light and fungal growth the PVDF body also has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and semisolids The PVDF outer junction sleeve...

Page 6: ...f old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup...

Page 7: ...coverable mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Meas...

Page 8: ...y indicator disappears 7 Place the tip of the electrode in the third buffer rinse beaker then in the third calibration buffer Wait until the measurement is stored and the stability indicator disappear...

Page 9: ...ace the sleeve Ensure the o ring is fixed inside the electrode Rinse off excess gel with distilled or deionized water Shake the probe down gently to eliminate any trapped air bubbles Soak the probe in...

Page 10: ...L HI721319 Soil auger RECOMMENDATIONS FOR USERS Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation intr...

Page 11: ...ion against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accord...

Page 12: ...rld Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810302 Printed in ROMANIA All manuals and user guides at all guides co...