Setup/Installation
19
Logged data on the edge
pH
® can be transferred from the meter to a PC by
following these simple directions. Suitable operating systems include Windows
(Xp minimum), OS X or Linux.
1.
Connect edge
pH
® to the PC using the supplied micro USB cable.
2.
Turn on edge
pH
®.
3.
Press
SETUP
and select “LOG ON EDGE”
4.
Press
MODIFY
then use
pq
keys to change to “EXPORT TO USB”
5.
Press
CFM
and the USB/PC Tag is displayed.
6.
Press
SETUP
to exit
The PC should detect the USB as a removable drive. Open the drive to view the
stored fi les. Log fi les are formatted as Comma separated values (*.CSV) and can
be opened with any text editor or spreadsheet program.
Note
:
•
Western Europe (ISO-88859-1) character set and English language are
suggested settings.
•
Other fi les may be visible depending upon computer settings. All fi les
stored will appear in this folder.
•
Adjust Font or column width appropriately. Adjust the decimal places if the
pH was logged with 0.001 resolution.
Interval logs are designated as pH or mV Lots. ie. PHLOT001, MVLOT002.
The Manual Lots are PHLOTMAN for pH and MVLOTMAN for ORP.
The Stability Lots are PHLOTSTAB for pH and MVLOTSTAB for ORP. All stability
logs, regardless of stability setting, are located in the same stability fi le for that
measurement.
Click on the desired log to view data.
Note
:
•
If “ºC !” appeared in log data, the electrode/probe was used beyond it’s
operation specifi cations and the data is not considered reliable.
•
If “ºC !!” appeared in log data, the temperature sensor within the probe or
electrode is broken and the device should be replaced. Logged data should
not be considered reliable.