•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

•  Fill the cuvette with 10 mL of unreacted sample 

and replace the cap. Place the cuvette into the 
meter and close the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking the meter 

is zeroed.

•  Remove the cuvette from the meter and unscrew 

the cap. Add the content of one packet of 

HI721-0

 reagent. Replace the cap and shake 

gently for 2 minutes until the powder is 
completely dissolved. Place the cuvette back 
into the meter.

•  Press and hold the button until the timer is 

displayed on the LCD (the display will show 
the countdown prior to the measurement) or, 
alternatively, wait for 3 minutes and press the 
button.

•  The instrument directly displays the 

concentration of iron in ppm. The meter 
automatically turns off after 2 minutes.

HI7

1-0

2

Measurement Procedure

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.00”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you use 
the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Functional Description

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Errors and Warnings

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument. For more information about Hanna 
Instruments and our products, visit www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us 
at [email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI721

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents for 6 tests
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0.00 to 5.00 ppm

Resolution

0.01 ppm

Accuracy

±0.04 ppm ±2% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Methods for the Examination of Water and 
Wastewater, 20

th

 edition, 3500-Fe B, Phenanthroline method. The re-

action between iron and reagent causes an orange tint in the sample.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 3 minutes of non-use and 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications