39
• Replace the cap tightly and shake the vials vigorously for about
30 seconds until all the powder is completely dissolved.
•
Insert the vials into the reactor and
heat them for 30 minutes at 105°C.
Note
: to obtain most accurate
results, it is strongly recommended
to remove the vials frome the reactor
after exactly 30 minutes.
•
At the end of the digestion period
switch off the reactor, place the vials
carefully in the test tube rack and
allow to cool to room temperature.
Warning
: as the vials are still hot, be
careful in handling them.
• Select the program number corresponding to Total Nitrogen LR on the
secondary LCD by pressing PROGRAM
t
and
s
.
•
Remove the cap from the vials and add the content of one packet
of Sodium Metabisulfite for Total Nitrogen analysis to each vial.
Replace the cap tightly and shake gently the vials for 15 seconds.
•
Wait for 3 minutes (without shaking the vials)
to allow the reaction to complete.
Nitrogen, Total LR
42
NITROGEN, TOTAL HIGH RANGE
SPECIFICATIONS
Range
10 to 150 mg/L
Resolution
1 mg/L
Precision
±3 @ 75 mg/L
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 420 nm
Method
Chromotropic acid method. A persulfate digestion
converts all forms of nitrogen to nitrate. Then the
reaction between nitrate and the reagents causes
a yellow tint in the sample.
REQUIRED REAGENTS
Code
Description
Q.ty/test Q.ty/set
*
Digestion Vial
1 vial
50 vials
---
Deionized Water
0.5 mL
1 bottle
---
Potassium Persulfate
1 packet
50 packets
---
Sodium Metabisulfite
1 packet
50 packets
HI 93767-0
Total Nitrogen Reagent 1 packet
50 packets
**
Reagent Vial
1 vial
50 vials
*
Digestion Vial identification
:
N
,
red
cap.
**
Reagent Vial identification
:
N
,
white
cap.
Note
: Store the unused vials in their container in a cool and dark place.
REAGENT SET
HI 93767B-50
Reagents for up to 49 tests. Contains:
Box 1: HI 93767B-50 Reagent Set
Box 2: HI 93767A&B-50 Reagent Set, for both
HI 93767A and HI 93767B parameter.
For other accessories see page 77.
REQUIRED ACCESSORIES
C 9800-01
Hanna reactor (115 VAC)
C 9800-02
Hanna reactor (230 VAC)
HI 740216
test tube cooling rack (25 holes)
HI 740217
Laboratory bench safety shield
For other accessories see page 77.
Nitrogen, Total HR