10
The sample and zero vial are inverted.
This indicates that the meter has lost its
configuration. Contact your dealer or the
nearest Hanna Customer Service Center.
ERROR MESSAGES
a) on zero reading:
The instrument cannot adjust the light level.
Please check that the vial does not contain
any debris.
There is not enough light to perform a
measurement. Please check the preparation
of the zero vial.
This indicates that the zeroing procedure
failed due to a low signal-to-noise ratio. In
this case press ZERO again.
There is too much light to perform a
measurement. Please check the preparation
of the zero vial.
The lamp is not working properly. Contact
your dealer or the nearest Hanna Customer
Service Center.
The lamp is not working properly. Contact
your dealer or the nearest Hanna Customer
Service Center.
b) on sample reading:
There is too much light for the sample
measurement. Please check if the right
sample vial is inserted.
71
Reagent Blank Correction
: This method needs a reagent blank
correction. A single blank vial may be used more than once; the blank
vial is stable up to one day (room temperature). For most accurate
measurement, run a blank for each set of measurements and always use
the same lot of reagents for blank and samples.
•
Preheat the Hanna Reactor C 9800 to 150 °C (302°F). For correct
use of the reactor follow Reactor Instruction Manual.
Use of the optional HI 740217 safety shield is strongly recommended.
Do not use an oven or microwave because leaking samples can
generate a corrosive and possibly explosive atmosphere.
• Remove the cap from two Reagent Vials.
• Add exactly 5.0 mL of sample to one vial (sample vial), and 5.0
mL of deionized water to the other vial (blank vial), while keeping
the vials at a 45-degree angle.
• Add the content of one packet of Potassium Persulfate for Phosphorus
analysis to each vial. Replace the cap tightly and shake gently the
vials until all the powder is completely dissolved.
•
Insert the vials into the reactor and
heat them for 30 minutes at 150°C.
Phosphorus, Total HR