WARRANTY

Instruction Manual

BL 983320-0
BL 983320-1

Panel-Mounted

EC Indicators &

Controllers

GENERAL DESCRIPTION

SPECIFICATIONS

Range

0.0 to 199.9 µS/cm

Resolution

0.1 µS/cm

Accuracy 

(@ 20°C/68°F)

±2% f.s.

Typical EMC Deviation

±2% f.s.

Probe

HI 7634-00

 EC/TDS probe (not included)

Temp.Compensation

Automatic from 5 to 50°C (41 to 122°F);

ß=2%/ºC

Calibration

Manual, through CAL trimmer

Dosing Contact

Maximum 2A (fuse protected), 250 Vac, 30 Vdc

Contact close when measure > setpoint

Setpoint

 Adjustable, from 0 to 199.9 µS/cm

Overtime

Adjustable, typically from 5 to approx. 30 minutes

Power Consumption

10 VA

Installation Category

II

Power supply

:

External (fuse protected)

BL983320-0

12 Vdc

BL983320-1

115/230 Vac ; 50/60Hz

Dimensions

83 x 53 x 99 mm (3.3x2.1x3.9”)

ACCESSORIES

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment
in which they are used. Operation of these instruments in residential areas could cause
unacceptable interferences to radio and TV equipment. Any variation introduced by the user
to the supplied equipment may degrade the instrument’s EMC performance. To avoid
electrical shock, do not use these instruments when voltages at the measurement surface
exceed 24 Vac or 60 Vdc. To avoid damages or burns, do not perform any measurement in
microwave ovens.

Remove the instrument from the packing material and
examine it carefully. If any damage has occurred during
shipment, immediately notify your Dealer or the nearest
Hanna Customer Service Center.
Each meter is supplied with:
• Mounting brackets
• Instruction manual
Note: Conserve all packing material until the instrument

has been observed to function correctly. Any defective
item must be returned in its original packing.

Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for the correct operation of the meter. Please read it carefully
before using the instrument.
If you need additional technical information, do not hesitate
to e-mail us at 

[email protected]

.

These instruments are in compliance with the 

 directives.

PRELIMINARY EXAMINATION

BL983320-0

 and 

BL983320-1

 are conductivity indicators

and controllers with a relay output designed for simplicity of
use in a wide range of  applications.
Connections and wiring to probe, power supply and contacts
are made via the terminal blocks on the rear panel.
The probe is easy to clean and requires little maintenance.
Other features include: automatic temperature compensation
of readings, single point calibration, overtime control system,
multi-colour LED for indicating if the meter is in measure-
ment/dosing/alarm condition, possibility to set (Off-Auto-On
switch) dosing action mode.
Two models are available:

• BL 983320-0

 powered at 12 Vdc

• BL 983320-1

 powered at 115 or 230 Vac

HI 7634-00

EC/TDS probe

HI 70033P

84 µS/cm calibration solution, 20 mL
sachet (25 pcs)

HI 7033M

84 µS/cm calibration solution, 230 mL

HI 7033L

84 µS/cm calibration solution, 500 mL

HI 7061M

Probe cleaning solution, 230 mL bottle

HI 7061L

Probe cleaning solution, 500 mL bottle

HI 710005

12 Vdc power adapter, US plug

HI 710006

12 Vdc power adapter, European plug

HI 710012

12 Vdc power adapter, Australian plug

HI 710013

12 Vdc power adapter, South African plug

HI 710014

12 Vdc power adapter, UK plug

HI 731326

Calibration screwdriver (20 pcs)

HI 740146

Mounting brackets

CE DECLARATION OF CONFORMITY

ASSEMBLING VIEW

These instruments are warranted for two years against defects in work-
manship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed main-
tenance are not covered. If service is required, contact the dealer from
whom you purchased the instrument. If under warranty, report the model
number, date of purchase, serial number and the nature of the failure. If
the repair is not covered by the warranty, you will be notified of the
charges incurred. If the instrument is to be returned to Hanna Instruments,
first obtain a Returned Goods Authorization Number from the Customer
Service department and then send it with shipment costs prepaid. When
shipping any instrument, make sure it is properly packaged for complete
protection.

 

 

 

Summary of Contents for BL 983320-0

Page 1: ...Instruction Manual BL 983320 0 BL 983320 1 Panel Mounted EC Indicators Controllers ...

Page 2: ...l the instrument has been observed to function correctly Any defective item must be returned in its original packing Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct operation of the meter Please read it carefully before using the instrument If you need additional technical information do not hesitate to e mail us at t...

Page 3: ... terminal blocks on the rear panel The probe is easy to clean and requires little maintenance Other features include automatic temperature compensation of readings single point calibration overtime control system multi colour LED for indicating if the meter is in measure ment dosing alarm condition possibility to set Off Auto On switch dosing action mode Two models are available BL 983320 0 powere...

Page 4: ...ASSEMBLING VIEW ...

Page 5: ...dependingonsetpointvalue ON dosing always active 3 MEAS key to set the instrument to measurement mode 4 SET key to display and set the setpoint value 5 SET trimmer to adjust the setpoint value 6 CAL trimmer 7 3 colourLEDindicator Green meter in measurement mode Orange Yellow dosing in progress Red blinking indicates an alarm condition ...

Page 6: ...ct 5 Jumper for enabling jumper in or disabling jumper removed theovertimecontrol 6 Trimmerforovertimesetting typicallyfrom5to30minutes All external cables connected to the rear panel should end with cable lugs A circuit breaker rated 6A max must be connected in close proximity to the equip ment and in a position easy to reach by the operator for disconnection of the instrument and of all the devi...

Page 7: ...es When the set time is exceeded any dosing action stops the LED indicator on the front panel will blink Red and the LCD will show the TIMEOUT warning message To exit the overtime condition set the OFF Auto ON switch to OFF position and then to Auto again For disabling the overtime feature simply remove the jumper on the rear panel Note The overtime system works only if the OFF Auto ON switch is i...

Page 8: ...l screwdriver adjust the SET trimmer until the desired setpoint value is displayed After 1 minute the meter automatically returns to the normal mode or press the MEAS key PROBE MAINTENANCE To improve the probe performance and prolong its life it is recommended to clean it regularly Immerse the tip of the probe in HI 7061 Cleaning Solution at least for one hour If a more thorough cleaning is requir...

Page 9: ... ß 2 ºC Calibration Manual through CAL trimmer Dosing Contact Maximum 2A fuse protected 250 Vac 30 Vdc Contact close when measure setpoint Setpoint Adjustable from 0 to 199 9 µS cm Overtime Adjustable typically from 5 to approx 30 minutes Power Consumption 10 VA Installation Category II Power supply External fuse protected BL983320 0 12 Vdc BL983320 1 115 230 Vac 50 60Hz Dimensions 83 x 53 x 99 mm...

Page 10: ... mL HI 7061M Probe cleaning solution 230 mL bottle HI 7061L Probe cleaning solution 500 mL bottle HI 710005 12 Vdc power adapter US plug HI 710006 12 Vdc power adapter European plug HI 710012 12 Vdc power adapter Australian plug HI 710013 12 Vdc power adapter South African plug HI 710014 12 Vdc power adapter UK plug HI 731326 Calibration screwdriver 20 pcs HI 740146 Mounting brackets ...

Page 11: ...anship and materials when used for their intended purpose and maintained according to instructions Probes are warranted for six months This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed main tenance are not covered If service is required contact the dealer from whom you purchased the instrument If under warranty report th...

Page 12: ...CE DECLARATION OF CONFORMITY ...

Reviews: