Operating Manual

Ultra 5.0

Ultra Centrifuge

Date of Purchase

Serial No.

Place of purchase

Summary of Contents for Ultra 5.0

Page 1: ...Operating Manual Ultra 5 0 Ultra Centrifuge Date of Purchase Serial No Place of purchase ...

Page 2: ...Copyright 2017 Hanil ScientificInc All rights reserved Contact Us If you have and questions contact Hanil ScientificInc or place of purchase sales ihanil com techsupport ihanil com ...

Page 3: ...on panel 9 4 2 Opening and closing the lid 10 4 3 Loading and removal of the rotor 10 4 4 Loading tubes 11 4 5 Setting RPM RCF 12 4 6 Setting run time 12 1 4 Electric safety information 4 5 Maintenance 6 Troubleshooting 6 1 General errors 15 7 Rotor 3 5 7 9 3 2 2 Power connetion 8 4 8 Start the centrifugation run 12 4 9 End the centrifugation 12 4 10 Pulse function 13 4 10 Saving Recall a program ...

Page 4: ...le or explosive vapor can not be centrifuged 7 The balancing work of samples should done in advance before operation 8 If the centrifuge is contaminated by toxic or radioactive samples or infectiousblood samples remove contaminants completely and take needful actions such as ventilation or isolation of centrifuge 9 Before operation rotor and chamber should be dry 10 Do not attempt to slow or stop ...

Page 5: ...indicatingdanger and warning Mark indicatinga place in danger of electric shock 1 4 Electric safety information 1 Use a power cord only provided with equipment 2 Use sockets with a protective earth conductorand suitable power cord 3 Do not use an extension cord 4 Do not place anything on the powercable 5 Do not block vents 6 If you have the following emergencies shut off the power supply and unplu...

Page 6: ...anual Optional items Rotors Display The display shows time rpm temperature Keypad Time and speed can be set Door Door protects the inside of the chamber Power switch On off the centrifuge RPM measuring window Visual inspection for rotor stop or for a speed check using a tachometer 5 l www ihanil com ...

Page 7: ...ure range 10 to 40 Time control 100 hr continuous Program memory 100 Noise level 65 dB Vacuum system Yes Rotor identification Automatic Dimension W x D x H mm 710 x 1 000 x 1 260 Weight without rotor 450 kg Power requirement VA 5 0 kVA Power input V Hz 220V 60 Hz 110V 50 Hz optional Cat No UL 5 0 sales ihanil com techsupport ihanil com l 6 ...

Page 8: ... bent due to the weight of the rotor Goodventailation For air circulation and safety maintain a 30cm clearance envelope aroundthe centrifugewhile the rotor is spinning Constant temperature humidity Centrifuge equipped with the sensitive electronic software which is fragile with humidity and temperature Must avoid direct ray or heater and be put in the ambience of controlled temp and air Avoid the ...

Page 9: ...ts on the label attached to the device Use sockets with a protective earth conductorand provided powercord 1 Connect the power cable to the power socket on the right side of the product and plug the cord into the outlet 2 Switch the centrifuge on using the power switch on the right side of the device sales ihanil com techsupport ihanil com l 8 ...

Page 10: ...on can be confirmed CALL Calling up a program SAVE Save a program VIEW SETTINGS Check the set value during centrifugation CLEAR erase the wrong number ENTER To set values ACCEL Set acceleration rates in 10steps DECEL Set deceleration rates in 10steps Reset The device reboots Lid Open centrifuge lid START Start centrifugation STOP Stop centrifugation Pulse After reaching the setting rpm it decelera...

Page 11: ...or Only use rotors parts and accessoriesprovided by Hanil ScientificInc Do not use scatchedor cracked rotor in use Loading Clean the motor shaft and the chamber When loading the rotor the temperature should be 10 to 30 degrees Load the rotor vertically onto the motor shaft Insert the wrench into the center hole of the rotor lid and turn it clockwise Check the rotor to make sure it loaded firmly Un...

Page 12: ... loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales Check the maximum load for each tube Put tubes into rotor holes Tubes located opposite each other must be the same type and contain the same quantity 11 l www ihanil com ...

Page 13: ...on to fix the value Press the Time button Press the numeric key to set centrifugationtime Press the Enter button to fix the value Press the Accel key or Decel key Press the Numeric keys to select the Acc Dec step The higher the setting step the faster the deceleration speed Decelstep 0 meansnaturalbrake 4 8 Start the centrifugation run Check the all parameters are correct and the lid is closed Pre...

Page 14: ...t the centrifugationtime speed Accel Decelvalues Press Save button You can store 100 programs on the device 0 99 Select the program slot by pressing the numeric keys and press the Enter button Press the Setup Enter key to save the program Recall a programs Press the Call button Press the program slot by pressing the numeric keys Press the Enter button to call up a program 13 l www ihanil com ...

Page 15: ... Use the stopwatchto check that the time setting is correct If the device is contaminated use a mild cleaning fluid to clean Make sure do not scratch the surface of equipment when cleaning it Do not use a metal sponge If the device is rusted remove it with a mild detergent and wipe it with a dry cloth Chamber If the rotor chamber is not dry wipe moisture from the chamber with a dry cloth If the ro...

Page 16: ...ely Lid cannot be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Checkthe lid latch is not damaged Unusual noise and vibration Check the device whether it is installed on the hard and flat place Reload the rotor symmetrically Reload the tubes symmetrically Tighten the clamping of the rotor with wrench by turning clockwise 15 l www ihanil c...

Page 17: ...7 Rotor 7 1 Angle rotors sales ihanil com techsupport ihanil com l 16 ...

Page 18: ...MEMO ...

Page 19: ...MEMO ...

Page 20: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of KOREA T 02 3452 8965 info ihanil com www ihanil com ...