7.1 Rotor

7. Rotor

Rotor

Tube 
Capacity

Required
Adaptor

Bore Ø x L (mm)
Radius (mm)

Max. RPM(rpm)
Max. RCF (xg)

      

AE

-

0.2-32

 

     (PCR Strip Rotor)

Hole angle : ∠45°

Max. Capacity : 32 x 0.2 mL

Size (ø x H) : ø172 x 38.5 mm

0.2mL

0.2mL strips

-

6.5 x 17

55.5

12,500

9,852

  

AE

-

2.0-24 

Hole angle : ∠45°

Max. Capacity : 24 x 1.5/2.0 mL

Size (ø x H) : ø172 x 38.5 mm

Max. height for tube fit : 56 mm

Incl. a lid

A2.0MP-24 incl. plastic lid

1.5/2.0 mL

-

11 x 38

83.7

15,000

21,206

0.5mL

TR0.5

8 x 37

78.5

15,000

19,747

0.2mL

TR0.2

6 x 21

68.5

15,000

17,231

  

AE-5c-10

Hole angle : ∠50°

Max. Capacity : 5 x 5.0 mL

Size (ø x H) : ø167.2 x 57.4 mm

Incl. a lid

5.0mL

-

17 x 53.8

83.6

15,000

21,853

Summary of Contents for M15R

Page 1: ...Operation Manual M15R Micro Centifuge Where Purchased Date of Purchase...

Page 2: ...fic Inc If you have any questions contact our Technical Support Center 82 2 3452 8966 techsupport ihanil com www ihanil com The appearance or specifications of the device is subject to partial change...

Page 3: ...ng Closing the Door 9 3 5 Installing Removing the Rotor 10 3 6 Rotor ID 11 3 7 Loading Tubes 12 4 Operation 13 4 1 Control Panel 13 4 2 Setting RPM RCF 14 4 3 Setting Run Time 14 4 4 Setting Temperatu...

Page 4: ...be deviced 7 The balancing work of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids strong bases cesium salt or alkaline dete...

Page 5: ...lug the powercord from outlet and contact your place of purchase Unusual noises or smell from the equipment Damage or wear of a power cord Breakdown of circuit breaker fuse or safety device If you spi...

Page 6: ...22 2 Product Description 2 1 Structure Door RPM measuring window Chamber Ventilation slots Display control panel Power Socket Drain cap Power Swithch Device Power cable Operating manual T wrench 2 2...

Page 7: ...ity 10 x 5 0 mL ACC DEC steps 9 10 steps Acc time to max speed 15 sec Dec time from max speed 15 sec Temperature range 10 to 40 Fast cool function Yes Program memory 10 Noise level 60 dB Imbalance cut...

Page 8: ...t due to the weight of the rotor Good ventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Cen...

Page 9: ...vice Use sockets with a protective earth conductor and provided power cord If the door is not closed the device does not operate 1 Connect the power cable to the power socket on the right side of the...

Page 10: ...otor When installing the rotor the temperature of the rotor and motor shaft should be at room temperature 10 30 2 Insert the rotor you want to fit into the motor shaft 3 Hold the rotor with one hand a...

Page 11: ...be centrifuged within the maximum speed 1 Press the Setup Enter button 2 Press the left or right button to change to the rotor ID setting mode Rotor Id Setting mode Rotor flashes on the display 3 Pres...

Page 12: ...loaded symmetrically Do not exceed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales 1 Check the maximum load for e...

Page 13: ...e cooling the rotor and chamber Left Left button is used to select parameter Right Right button is used to select parameter Setup Enter Setup Enter button can be pressed to go into setup mode or compl...

Page 14: ...hes RCF setting mode RCF xg display flashes 3 Press Up or Down button until the desired RPM RCF is displayed 4 Press the Setup Enter button 1 Press the Setup Enter button 2 Press the Left or Right but...

Page 15: ...e display 00 flashes 3 Press Up or Down button until the desired temperature is displayed 1 Set the desired Temperature Refer to 4 4 Setting Temperature 2 Press the Down button more than 2sec A rotor...

Page 16: ...splay flashes 0 Decel rate setting mode Decel display flashes 0 3 Press the Setup Enter button 4 6 Setting Acceleration Deceleration Start 1 Check all parameters are correct 2 Press the Start Stop Pul...

Page 17: ...F Run time Accel Decel rate 2 Press the Left of Right button until Prog display flashes 0 You can save 10 0 9 programs 3 Press Up or Down button until the desired program number is displayed 4 Press t...

Page 18: ...ive chemicals on the device such as alchol benzene acetone or phenol 3 Make sure do not scratch the surface of deivice when cleaning it Do not use a metal sponge If the device is rusted remove it with...

Page 19: ...d Check the lid is closed completely Door cannot be opened Press the Door button Check the power connection Door cannot be closed Remove the dirt at the door latch and close the lid Check the lid latc...

Page 20: ...System error Contact the Technical support team E10 Inner Temperature Over OR Inner temperature sensor error Occurs when the set temperature is not reached Inner temperature sensor error Make sure th...

Page 21: ...ce door cannot be opened you can activte the emergency open manually 1 Wait for rotor to stop before activating the emergency open 2 Insert the T wrench into the emergency open hole and turn it clockw...

Page 22: ...0 2mL strips 6 5 x 17 55 5 12 500 9 852 AE 2 0 24 Hole angle 45 Max Capacity 24 x 1 5 2 0 mL Size x H 172 x 38 5 mm Max height for tube fit 56 mm Incl a lid A2 0MP 24 incl plastic lid 1 5 2 0 mL 11 x...