1. General Considerations

1.1 Safety

Follow precautions and all the safety requirements described on this user manual 

to prevent any damage and failure of equipment and loss of lives.

1. The centrifuge should be installed on flat surface to maintain level.

2. Check the voltage to be used, before connecting the centrifuge to the power source.

3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc.

Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by 

using parts and accessories not recommended.

4. Do not exceed the maximum rated speed of the rotor or buckets in use.

5. Make sure to prepare necessary safety measures before using samples that are toxic or 

radioactive samples or pathogenic or samples or infectious blood.

6. Substances that may generate volatile or explosive vapor can not be centrifuged.

7. The balancing work of samples should done in advance before operation.

8. To ensure safe use of the device, do not expose the device to strong acids, strong bases, 

cesium, salt, or alkaline detergents.

8. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood 

samples, remove contaminants completely and take needful actions such as ventilation 

or isolation of centrifuge. 

9. Before operation, rotor and chamber should be dry.

10. Do not attempt to slow or stop the spinning rotor by hand.

11. Only centrifuge with rotor and rotor lid firmly tightened.

12. Do not block vents.

13. When serving the centrifuge, be sure to remove contaminants in advance.

14. Please contact the place of purchase or Hanil Scientific Inc. for product repairs.

15. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the 

centrifuge while the rotor is spinning.

16. Turn the power switch off after using the device.

17. Unplug the power plug before cleaning or left unused for a long period of time.

Summary of Contents for Fleta5

Page 1: ...Operating Manual FLETA 5 Clinical Centrifuge Date of Purchase Serial No Place of purchase...

Page 2: ...se 82 2 3472 0727 Inquiry info ihanil com Order sales ihanil com Tech support techsupport ihanil com The contents in this operating manual are subject to change for device improvement Copyright 2018 H...

Page 3: ...id 11 4 3 Loading and removal of the rotor 11 4 4 Setting rotor bucket ID 12 4 5 Loading tubes 13 4 6 Setting RPM 14 1 4 Electric safety information 4 5 Maintenance 6 Troubleshooting 6 1 General error...

Page 4: ...atile or explosive vapor can not be centrifuged 7 The balancing work of samples should done in advance before operation 8 To ensure safe use of the device do not expose the device to strong acids stro...

Page 5: ...lid Mark indicating danger and warning Mark indicating a place in danger of electric shock 1 4 Electric safety information 1 Use a power cord only provided with equipment 2 Use sockets with a protecti...

Page 6: ...or is loaded Operating panel Set desired run conditions Power switch On off the centrifuge Emergency lid open tool Active the emergency release on the bottom of the device Fuse case If the wiring is s...

Page 7: ...2 Product description 2 2 Delivery package Mainbody Power Cable Operating manual Twrench for rotor locking Optional items 6 l www ihanil com...

Page 8: ...x speed 20 80 sec Dec time from max speed 26 82 sec Time control 100 hr continuous Program memory 100 Noise level 65 dB Imbalance cutoff Yes Dimension W x D x H mm 570 x 475 x 325 Weight without rotor...

Page 9: ...ent due to the weight of the rotor Goodventailation For air circulation and safety maintain a 30cm clearance envelope around the centrifuge while the rotor is spinning Constant temperature humidity Ce...

Page 10: ...rical requirements on the label attached to the device Use sockets with a protective earth conductor and provided powercord 1 Connect the power cable to the power socket on the back side of the produc...

Page 11: ...ayed Display Time Centrifugation time is displayed Set to 0 0 0 for continuous centrifugation Display Prog Program No is displayed Display Acc Acceleration rate is displayed Display Dec Deceleraterate...

Page 12: ...ccessories provided by Hanil Scientific Inc Do not use scatched or cracked rotor in use Loading Clean the motor shaft and the chamber When loading the rotor the temperature should be 10 to 30 degrees...

Page 13: ...e knob to move the cursor to rotor ID mode 3 Turn the Enter knob until the desired rotor bucket ID displayed 4 Press the Enter button 5 Do you save appears Turn the mode knob to select YES 6 Press the...

Page 14: ...eed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales Check the maximum load for each tube Put tubes into rotor hol...

Page 15: ...5 Do you save appears Turn the Mode knob to select Yes 6 Press the Enter knob 4 7 Setting RCF 1 Press the Mode knob in stand by mode 2 Press the Enter knob 3 Turn the Mode knob to select RCF setting m...

Page 16: ...5 Do you save appears Turn the Mode knob to select Yes 6 Press the Enter knob 4 9 Setting Acceleration Rate 1 Press the Enter knob 2 Turn the Mode knob to select ACC setting mode ACC setting mode ACC...

Page 17: ...r knob to save the value 5 Do you save appears Turn the Mode knob to select Yes 6 Press the Enter knob 4 11 Saving a Program 1 Press the Enter knob until desired program number is displayed You can se...

Page 18: ...isplayed You can see that the program number is changed on the PROG display 4 13 Starting or Stopping a Run Starting a run 1 Check that all parameters are correct and the lid is closed 2 Press the Sta...

Page 19: ...e a mild cleaning fluid to clean Make sure do not scratch the surface of equipment when cleaning it Do not use a metal sponge If the device is rusted remove it with a mild detergent and wipe it with a...

Page 20: ...t be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Checkthe lid latch is not damaged Unusual noise and vibration Check the device whether it...

Page 21: ...to operate again please call please call Field Service Engineer E3 Motor overheat Motor overheat Shut off the power supply and then turn on the power switch again to check the device E4 Motor operatio...

Page 22: ...e E13 Low voltage error Low Voltage Contact Hanil Scientific Inc E15 RPM Senser error RPM Sensor error Rotor identification error Check if the id disc is on the bottom of the rotor E17 Communicati on...

Page 23: ...active the emergency release 6 Troubleshooting 6 3 Emergency lid release Turn off the power switch Wait for the rotor to stop completely Find the emergency release key at the bottom the decvice Pull d...

Page 24: ...7 Rotor 7 1 Rotors and adaptors Recommended RPM 4 500 info ihanil com techsupport ihanil com l 23...