6. Troubleshooting
6.2 Error code
Error
code
Error
message
Cause
Recommended action
E1
Imbalance
error
- When an imbalance is
detected
- Check weight-balances of samples and then
turn off and on the device for checking.
E2
Over speed
error
- over speed than set speed
- Shut off the power supply, and then, turn on
the power switch again to check the device.
- If the error code shows continuously although
you try to operate again, please call please call
Field Service Engineer.
E3
Motor
overheat
- Motor overheat
- Shut off the power supply, and then, turn on
the power switch again to check the device.
E4
Motor
operation
error
-Motor operation error
- Check that the rotor is rotating through the
RPM measuring window
- Shut off the power supply, and then, turn on
the power switch again to check the device.
E5
Lid
open(Operat
ng)
- Lid open during operation
- Shut off the power supply, and contact Hanil
Scientific Inc.
E7
System error
- System error
- Contact Hanil Scientific Inc.
E11
Motor
temperature
sensor error
Motor temperature sensor
error
- Shut off the power supply, and then, turn on
the power switch again to check the device.