Check the cushion rubbers in the bucket
(HBT15-8) before use. When reinserting, must
insert it in the correct direction.
Spread the grease lightly on the bucket where it
contacts the rotor and the pivot.
Check the Cushion Rubbers
Maintenance of Swing-Out Rotor
HSM-2
MTP
Cell Culture
DWP
Max.2k (Rotor ID. 4)
HBTM
/ Max. 3k (Rotor ID. 3)
HBT50-1
1 x 50 ml / Max. 5.5k (Rotor ID. 2)
HBT50-2
2 x 50 ml / Max. 5.5k (Rotor ID. 2)
HBT50c-1
1 x 50 ml conical / Max. 5.5k (Rotor ID. 6)
HBT15-8
8 x 15 ml / Max. 4k (Rotor ID. 1)
HBT15c-4
4 x 15 ml conical / Max. 4k (Rotor ID. 5)
Bucket
Rotor
If the centrifuge lid cannot be opened, find the “emergency
release key” at the bottom of the device. And pull down the
“emergency release key”. (Before activating emergency
open, switch off the device and wait for the rotor to stop.)
Allow a safety clearance of 30 cm around the
centrifuge during operation.
Allow a Safety Clearance of 30 cm
30 cm
30 cm
30 cm
Quick Guide
Fleta 5
Cat. No. : FL-5
Emergency Lid Open
700VA
220~240V
50/60Hz
HSR-4S
Rev.0
Proudly made by Jiyoung Shim, Hyewon Jeong
Order & Inquiry l +82-2-3452-9996 l [email protected] l www.ihanil.com
Service l +82-2-3452-8966 l [email protected]
After relocating the device or loading
the rotor, balance them by using
a leveler aligner.If necessary, level the
device by adjusting the ‘Leveling foot’
on the bottom of the device.
Balancing the Centrifuge and Rotor
Check the rotor,buckets and tubes
are free of scratch and crack
before operation.
Check the Accessories for
Abnormalities
Check the rotor is firmly tightend
onto the motor shaft.
Check the Rotor Locking
Load tubes symmetrically into
the rotor with identical amount.
Load Tubes Symmetrically
Be
fo
re Ope
ration
After Ope
ration
Clean the chamber and rotor by wiping with a
cloth soaked in mild detergent and dry
completely.
Cleaning
Rotors & Bucket
Maintenance
Electrical Safety
Cautions for Use
Connect the device to voltage souces, which
correspond to the power requirements on the
label attached to the device.
Check the wall outlet is earth-grounded, surge-
protected and can supply enough power.
Check the Line Voltage
Grounded Outlet
Switch Off the Device
Switch off the device after use.
Unplug the power plug before cleaning or
unused for a long period of time.