Check the cushion rubbers in the bucket 
(HBT15-8) before use. When reinserting, must 
insert it in the correct direction.

Spread the grease lightly on the bucket where it 
contacts the rotor and the pivot.

Check the Cushion Rubbers 

Maintenance of Swing-Out Rotor

HSM-2

MTP  

 Cell Culture 

DWP

   Max.2k (Rotor ID. 4) 

HBTM

 / Max. 3k (Rotor ID. 3) 

HBT50-1

  1 x 50 ml / Max. 5.5k (Rotor ID. 2) 

HBT50-2

  2 x 50 ml / Max. 5.5k (Rotor ID. 2) 

HBT50c-1

  1 x 50 ml conical / Max. 5.5k (Rotor ID. 6) 

HBT15-8

  8 x 15 ml / Max. 4k (Rotor ID. 1) 

HBT15c-4

  4 x 15 ml conical / Max. 4k (Rotor ID. 5) 

Bucket

Rotor

If the centrifuge lid cannot be opened, find the “emergency 
release key” at the bottom of the device. And pull down the 
“emergency release key”. (Before activating emergency 
open, switch off the device and wait for the rotor to stop.)

Allow a safety clearance of 30 cm around the 
centrifuge during operation.

Allow a Safety Clearance of 30 cm

30 cm

30 cm

30 cm

Quick Guide

Fleta 5

Cat. No. : FL-5

Emergency Lid Open

700VA
220~240V 

50/60Hz 

HSR-4S

Rev.0

Proudly made by Jiyoung Shim, Hyewon Jeong 

Order & Inquiry l +82-2-3452-9996 l [email protected] l www.ihanil.com
Service  l +82-2-3452-8966 l [email protected]

After relocating the device or loading 
the rotor, balance them by using 
a leveler aligner.If necessary, level the 
device by adjusting the ‘Leveling foot’ 
on the bottom of the device.

Balancing the Centrifuge and Rotor

Check the rotor,buckets and tubes 
are free of scratch and crack 
before operation.

Check the Accessories for 
Abnormalities

Check the rotor is firmly tightend 
onto  the motor shaft.

Check the Rotor Locking

Load tubes symmetrically into 
the rotor with identical amount.

Load Tubes Symmetrically

Be

fo

re Ope

ration

After Ope

ration

Clean the chamber and rotor by wiping with a 
cloth soaked in mild detergent and dry
completely.

Cleaning

Rotors & Bucket

Maintenance

Electrical Safety

Cautions for Use

Connect the device to voltage souces, which 
correspond to the power requirements on the 
label attached to the device.

Check the wall outlet is earth-grounded, surge- 
protected and can supply enough power.

Check the Line Voltage

Grounded Outlet

Switch Off the Device

Switch off the device after use.
Unplug the power plug before cleaning or 
unused for a long period of time.