Follow precautions and all the safety requirements described on this user manual 
to prevent any damage and failure of equipment and loss of lives.

1. The centrifuge should be installed on flat surface to maintain level.

Abnormal vibration is caused when the device is installed on an inclined surface.

2. Check the voltage to be used, before connecting the centrifuge to the power source.

3. Only use rotors, parts, and accessories provided by Hanil Science Inc.

Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by 
using parts and accessories not recommended.

4. Do not exceed the maximum rated speed of the rotor or buckets in use.

5. Make sure to prepare necessary safety measures before using samples that are toxic or 

radioactive samples or pathogenic or samples or infectious blood.

7. Substances that may generate volatile or explosive vapor can not be centrifuged.

8. The balancing work of samples shoud be done in advance before operation.

6. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples,

remove contaminants completely and take needful actions such as ventilation or isolation 
of centrifuge. 

9. Before operation, rotor and chamber should be dry.

10. Do not attempt to slow or stop the spinning rotor by hand.

11. Only centrifuge with rotor and rotor lid firmly tightened.

12. Do not block vents.

13. When serving the centrifuge, be sure to remove contaminants in advance.

16. Please contact the place of purchase or Hanil Scientific Inc. for product repairs.

15. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the 

centrifuge while the rotor is spinning.

14. Repairs must be carried out by a technologist authorized byHanil scientific Inc.

[email protected] / [email protected]      l 3

1. General Considerations

1.1 Safety

Summary of Contents for Fleta 5

Page 1: ...Operating Manual FLETA 5 Clinical Centrifuge Date of Purchase Serial No Place of purchase...

Page 2: ......

Page 3: ...Opening and closing the lid 11 4 3 Loading and removal of the rotor 11 4 4 Loading tubes 12 4 5 Setting RPM RCF 13 4 6 Setting run time 13 1 4 Electric safety information 4 5 Maintenance 6 Troubleshoo...

Page 4: ...ectious blood 7 Substances that may generate volatile or explosive vapor can not be centrifuged 8 The balancing work of samples shoud be done in advance before operation 6 If the centrifuge is contami...

Page 5: ...he lid Mark indicatingdanger and warning Mark indicatinga place in danger of electric shock 1 4 Electric safety information 1 Use a power cord only provided with equipment 2 Use sockets with a protect...

Page 6: ...Where the rotor is loaded Operating panel Set desired run conditions Power switch On off the centrifuge Emergencydoor open tool Active the emergency release Fuse case If the wiring is short circuited...

Page 7: ...2 Product description 2 2 Delivery package Fleta5 Mainbody incl HBT15 8 Power Cable Operating manual Twrench for rotor locking Optional items 6 l www ihanil com...

Page 8: ...ax speed 20 80 sec Dec time from max speed 26 82 sec Time control 10 hr continuous Program memory 100 Noise level 65 dB Imbalance cutoff Yes Dimension W x D x H mm 570 x 475 x 325 Weight without rotor...

Page 9: ...bent due to the weight of the rotor Goodventailation For air circulation and safety maintain a 30cm clearance envelope aroundthe centrifugewhile the rotor is spinning Constant temperature humidity Cen...

Page 10: ...ts on the label attached to the device Use sockets with a protective earth conductorand provided powercord 1 Connect the power cable to the power socket on the right side of the product and plug the c...

Page 11: ...yed Display Time Centrifugation time is displayed Set to 0 0 0 for continuous centrifugation Display Prog Program No is displayed Display Acc Acceleration rate is displayed Display Dec Deceleraterate...

Page 12: ...use rotors parts and accessoriesprovided by Hanil ScientificInc Do not use scatchedor cracked rotor in use Loading Clean the motor shaft and the chamber When loading the rotor the temperature should...

Page 13: ...ly Do not exceed the maximum rated speed of the tube Same volume of sample should be put on tubes Check symmetric loading by balncing tubes with scales Check the maximum load for each tube Put tubes i...

Page 14: ...eft or right to input the value Press the Enter button to save the value Do you save Yes No Turn Mode button to select Yes or No Press the Enter button to finish setting 4 6 Setting run time Press the...

Page 15: ...step 9 step is the fastest step Recommend accel decal5 step 4 8 Radius input When the lid is closed the rotor is recognized automatically After recognizing the rotor the radius is automatically set T...

Page 16: ...speed Accel Decel values Press Enter button Turn Enter button to set Program No Turn Move button to select YES Press Enter button to save a program Saving a program Recall a programs Turn Enter butto...

Page 17: ...a mild cleaning fluid to clean Make sure do not scratch the surface of equipment when cleaning it Do not use a metal sponge If the device is rusted remove it with a mild detergent and wipe it with a...

Page 18: ...t be opened Press the Lid open button Lid cannot be closed Remove the dirt at the door latch and close the lid Checkthe lid latch is not damaged Unusual noise and vibration Check the device whether it...

Page 19: ...e lid by emergency key Not close the lid perfectly To close the door No working lid pin Lid is not closed tightly Press the Door button to open the lid and then close it again Turn off the main power...

Page 20: ...ergency release 6 Troubleshooting 6 3 Emergency lid release Turn off the power switch Wait for the rotor to stop completely Find the emergency release key at the bottom the decvice Pull down the emerg...

Page 21: ...7 Rotor 7 1 Rotors and adaptors 20 www ihanil com...

Page 22: ...Hanil Scientific Inc 16 Arayukro Gimpo 10136 Rep of KOREA T 02 3452 8965 info ihanil com www ihanil com...

Reviews: