Error
Not open lid
No display
None
- No power connection.
-
Check the power cable
.
-
Check the main fuse
in laboratory and device
Display
- Failure of power supply
Cause
Solution
Unknow Rotor ID!
- No rotor
- Load a rotor.
- Turn off the main power and
turn on again.
- Remove foreign matter from
the rotor or sensor.
-
Errors in drive or rotor
recognition function.
- Foreign substances in
the rotor or sensor.
No running
None
-Fauiure of power supply
- Stop the running and open
the lid by emergency key.
Not close the lid
perfectly.
To close the door
- No working lid pin
- Lid is not closed tightly.
- Press the Door button to open
the lid and then close it again.
- Turn off the main power and
turn on again.
** Imbalance
Error! **
During
Acceleration,
the device
vibrates and
switch off.
- Not load symmetrically
- Not tighten the rotor with shaft
- The device is placed uneven place
- Check the rotor and samples
- Tighten the rotor properly.
- Place the device on flat place
Failure of power
** Power
Failure! **
- Failure of power supply
- Check a outlet
** Motor
Overheat! **
- Motor over heat
- Stop the centrifugation and
lower the temperature
** Over Speed! ** - Run above 105% of the set spped
- Stop and run again
Etc.