System Fault Warning
This pop-up window appears if the machine detects a
communications error between its two processors. This
should be a very rare event if it happens at all, and may
not recur if the mains power is switched off and on again.
If it persists, then contact ACP Customer Support at (1-
800-350-1100).
Testing the Output
Enter the manual operation screen and select the 4ΔT mode. Touch the “start” button to initiate output to the applicator
head. Place the head in an upward facing position the power meter should indicate detected power level of 200watts.
Use the tuning knob on the head to maximize forward power. If it does not, then qualified service personnel should be
called.
Note:
Allow the applicator to warm up for 2 minutes for proper readings.
Output Tuning Bar
A tuning bar feature exists on the device. However, the
appearance of the tuning bar on the treatment run screen,
is optional.
To turn off the tuning bar, enter the Systems Settings
menu from the Home screen. Once in the Systems
Settings menu press Device Output Tuning.
Press tuning bar ON and it will change to OFF. The tuning bar will no longer appear on the run screen.
Using Tuning Bar
Place the head in an upward facing position toward the ceiling.
Enter the systems settings menu from the Main Menu screen. Once in the Systems Settings menu press Device
Output Tuning.
To optimize the energy output: Press the START button below. Gradually turn the Tuning Knob on the applicator
head to display the most amount of green and the least amount of yellow on the bar below. Press DONE when
complete
.