Chart 3.3 the main measuring interface
In the data display interface, sampling period can be shown as well
as the working circuit temperature, current, voltage, and ISO4406
NAS1638 corresponding standard level. Users can observe the
current different standards according to the need of test information.
Sampling period is 6 seconds (default). The laser parameters such
as current, receiving voltage are sensor diagnostic parameters.
2.4 Instrument Rinsing
Hose flushing is recommended after each operation. Petroleum
ether is the recommended fluid for flushing. Insert the oil-in hose in
flushing fluid and the oil-out hose into a waste container or bottle.
Select the "rinsing" button to start the pump. Keep rinsing liquid
flowing for a minute and then click the "stop" button. Remove the
hoses, let the pump run for a few minutes to eliminate the
remaining rinsing liquid in hoses. Flushing now is completed.
Note: For hose rinsing, we recommend that high-speed rinsing is
adopted in the beginning, then switch it to low-speed.
2.5 Battery Recharging
Battery is accessories. If you buy the instrument with battery and in
the situation of insufficient power, the display panel owe electric
lights will be on. please recharge in time. Firstly, connect the