the operator can perform by themselves. For other problems,
please contact our engineers for further assistance.
5 Notes
Please select the appropriate product model that meets your
actual requirement. Low Pressure model: Not less than 20 psi
(1.3 Bar) of pressure difference. Please make sure there is no
large particles greater than 100µm (visible by naked eyes) or
transparent floccules, as the big particles can clog up at the
sensor.
Please make sure there are no obvious bubbles in the liquid,
otherwise the accuracy of measurement will be affected.
The sample volume should be no less than 100ml.
this product is a precision laser instruments. Strong vibration will
Simple Fault
Resolution for Fault
Unable to start the instrument
Check if the power cable is plugged in and
power turned on.
Unable to print
Check to see if the printer is short of paper.
No oil comes out from the oil-out
Check to see if the oil in hose is connected;
Check to see if pressure is within the required
parameters.
Not less than 20 psi (1.3 Bar) of pressure
difference);
Possible blockage in sensor.(needs to be sent
back to supplier for resolution)
Pop-up window
indicating
"errors".
Unable particle count.
1.Too many bubbles in oil. Eliminate air by
using ultrasonic bath.
2. Re-enter the measuring interface and try
again.