MFI 4050
V1.3
32
Replace all existing equipment with the newly supplied calibrated parts.
Changing the inner barrel
The inner barrel can be removed using the removal tool. If the inner barrel is
stuck in the instrument try heating the barrel. Once it is loose leave to cool before
withdrawing it from the instrument.
IF REMOVING THE BARREL HOT WEAR HEAT PROOF GLOVES
Fit the new inner barrel using the tool.
Temperature Verification and Calibration
The new calibration probe supplied with the calibration Pack has been certified
by a UKAS calibration laboratory this piece of calibration equipment is used to
calibrate the MFI with the new inner barrel fitted.
Power up the instrument
Press calibrate
Enter the serial number of the calibration pack.
Enter the calibration code for the calibration pack.