MFI 4050
V1.3
31
When the MFI instrument is not in use it should be powered down. This will leave
the instrument in a safe state with the test mass lowered. Remove the test piston
or the cleaning probe if fitted to the test weight. In the system menu press the off
button.
Calibration
The Hanatek MFI is supplied calibrated. We recommend re-certifying the
instrument annually. This can be done by a Hanatek approved calibration
engineer or the user can self certify the instrument using a Hanatek Calibration
pack.
The calibration pack will enable the user to calibrate the instrument using UKAS
certified calibration artefacts.
The re-calibration kit consists of the following certified parts:
Calibration Probe (replaces existing probe)
Standard Testing Die (replaces existing Die)
Standard Testing Piston (replaces existing Piston)
Replacement inner Barrel (replaces existing Inner Barrel)
Cleaning Patches
Barrel Removal Tool
Step 1