C8484-16C Instruction manual Ver.1.5
24
(2) 2×2 binning readout
Exposure and 2
2 binning readout repeat according to the sequence set internally.
Command
: AMD E (External control mode)
External control pulse : ATP N {ATP P}
Scan mode
: SMD S (Binning)
Exposure time
: EMD E,EST n (Internal setting)
Binning
: SPX 2 (2×2)
1V (vertical)
: 87.4 µs
Setting count
: 1H to 12 930H (22.65 µs to 1.129 s)
Minimum Repeat cycle can be calculated using the formula below.
Minimum repeat cycle = 46.9 ms (readout time) + set exposure time
Exposure time can be calculated using the formula below.
Exposure time = 22.65 µs + (n-1) × 87.4 µs
(n = 1 to 12 930)
External trigger pulse
min. 1 µs
CCD exposure
Exposure time = EST n
Electrical shutter signal
max. 2.2 µs
Readout signal
from CCD chip
Digital data output
Readout
period