qCMOS Camera C15550-20UP Instruction manual_Ver.1.0
22
8-2 CONNECTION OF WATER COOLING HOSES
Figure 8-3
1. Place the camera on the stable table.
2. Connect water cooling hose into the WATER connector on the camera.
- Insert the hose fully into the WATER connector on the camera. (as shown in Figure 8-3)
- Confirm the hose stops at it.
3. Set the camera onto a microscope (If the camera is used on the microscope).
If it is easy to connect the hose onto the camera after the camera is set onto the microscope then it is
OK to connect the hose after the camera is set on the microscope.
4. Connect the hose onto the circulating water cooler.
Follow the instruction on the circulating water cooler when you connect the hose onto the circulating
water cooler.
5. Turn on the circulating water cooler and confirm the cooling water is flowing normally.
•
Stop the circulating water cooler when the water flow is abnormal or water
drop or splash is found.
Water cooling hose
WATER
connector