6. NAME AND FUNCTION OF PARTS
Figure 6-1
•
Place the camera the ventilation ports to be lateral side. Do not place the
rear panel of the camera, which connectors are located, to be at the
bottom (Do not block ventilation openings.).
Lens mount
C15550-20UP can be attached to C-mount lens or an optics system.
Note
•
The depth of the C-mount is 7 mm. Screwing in the C-mount too deeply might
scratch the glass surface.
WATER connector [WATER] (when using Water-cooling)
It connects the camera and the circulating water cooler with the cooling water hoses. The connector
position of WATER IN/OUT is not specified.
•
See 8 “WATER COOLING” for instruction of water-cooling.