Mini-spectrometers 

KXX-A12040  A 

This document describes general precautions for using Hamamatsu mini-spectrometers. 
 

If the product comes with special precautions on the delivery specification sheet, then be sure to strictly 
comply with those instructions. 

 

  1. All models   

 

Mini-spectrometers (hereafter “the product”) are high precision optical components. Do not apply excessive 
vibration and shock to the product. Avoid operating and storing it in dusty locations or in areas exposed to 
water, high temperature or high humidity. 

 

When cleaning the enclosure, wipe it with a clean, soft, dry cloth. Do not use organic solvents such as 
thinner and acetone. 

 

Do  not  apply  excessive  force  to  the  connector.  Doing  so  may  cause  a  poor  connection  or  damage  the 
connector. 

 

Use the product at incident light levels that do not cause the A/D count of each pixel to become saturated. 
Correct measurement may not be possible when some pixels become saturated. Set a shorter integration 
time if the A/D count is saturated. If the integration time cannot be shortened, we recommend you use a 
neutral density filter. 

 

Even when no light is input, the product generates an output consisting of an offset component and dark 
signal. The offset component is not dependent on integration time, but the dark signal is dependent on it. 

 

Coefficients  for  converting  image  sensor  pixel  numbers  into  wavelengths  are  described  in  the  final 
inspection sheet that comes with each product. For module types, those coefficients are also written into the 
EEPROM in the mini-spectrometer. However, we do not provide coefficients for converting the output value 
of each pixel to a value proportional to the incident light level. 

 

  2. Module type   

 

The optical connector is exposed to outside air, so when not using the product for extended periods, place a 
cap on the optical connector or connect the input optical fiber to the connector. When connecting the optical 
fiber, clean the ferrule on the optical fiber as needed. 

 

Power supplied through a USB port from a PC is limited to 5 V/500 mA per port due to USB specifications. So 
power exceeding this limit cannot be supplied to the product. If connecting two or more products to one 
USB port via a hub, use a hub with a power supply. 

 

Evaluation software differs according to the product model. Be sure to install the evaluation software that 
supports the product you will be using. 

Precautions 

Summary of Contents for C12666MA

Page 1: ...when no light is input the product generates an output consisting of an offset component and dark signal The offset component is not dependent on integration time but the dark signal is dependent on...

Page 2: ...e supplied with it Use caution when turning on the power supply because an inrush current flows at that time Table 1 shows the maximum current in steady state and typical inrush current Table 1 Maximu...

Page 3: ...h a wrist strap having protective resistance of 750 k to 35 M directly to the skin and ground the strap If the wrist strap does not include protective resistance there is a risk of electric shock haza...

Page 4: ...ductive case Avoid using plastic or styrofoam as they may generate static electricity by vibration during shipping etc causing breakdown or deterioration of the heed type Use a conductive carrying cas...

Page 5: ...fluctuations so separate it as far as possible from the video signal line and voltage supply line Apply the input clock pulses at the specific timing and voltage to the product connection terminals on...

Page 6: ...the optical fiber SMA connector and then use the slotted holes to secure the mini spectrometer at a position where the optical fiber is free from stress Figure 3 Precautions when connecting input opt...

Page 7: ...cure it with resin or a holder etc 3 Precautions for designing the circuit board The packaging of the micro spectrometer is electrically conductive so be careful when designing the circuit to avoid sh...

Page 8: ...d of time seal it in a moisture proof package containing silica gel or keep it in a low humidity desiccator 4 Input optical fibers option When connecting the input optical fiber clean the ferrule on t...

Reviews: