C11440-22CU/C11440-22CU01 Instruction manual_Ver.1.4
8-3 DISCONNECTION OF WATER COOLING HOSES
•
Remove the water cooling hoses only when it is necessary to remove.
(1) Turn off the camera power and all peripheral devices including circulating water cooler.
(2) Remove the hose on circulating water cooler side.
Follow the instruction on the circulating water cooler when you disconnect the hose
from the circulating water cooler.
(3) Remove water or water drop inside the hose and the camera by air.
• Blow Air from one side of hose. Prepare water absorption sheet (such as Waste,
Towel or so) and catch pan on another side of hose in order to avoid water drop or
water splash.
• Blow Air until no water drop come out.
(4) Remove the camera from the microscope (if the camera is used on the microscope).
It is not necessary to remove the camera from the microscope if it is possible to remove
the hoses from the camera as it is.
(5) Place the camera on the stable table.
Put the lens cap on to protect the sensor.
(6) Change the WATER connector direction to be downward.
Prepare water absorption sheet (such as Waste, Towel or so) and catch pan.
(7) Remove hoses one by one, and wipe water.
Disconnect hoses with pushing button while being careful not to splash water.
Figure 8-4
•
Cooling water may be left inside the camera even after hoses are removed.
In such case, remove water inside by blowing air from connectors. Be careful
not to splash water onto the camera.
Button
21