Appendix-B
35
Digital radiography system, dental C10819 Series
Question
Continuous imaging is not possible
Answer
Is continuous imaging performed with short accumulation time?
When the image data is large, the data processing fails to meet the stipulated
deadline and the imaging is not performed. Perform the imaging by increasing
the accumulation time.