Appendix-B
34
Digital radiography system, dental C10819 Series
Question
Attached application software is not running.
Answer
Is the power being supplied to the main unit?
Please verify that the LED of the main unit is on.
Question
Unable to create image
Answer
1) Is the sensor unit set in disconnected status (Idle/Sleep mode)?
Set the sensor unit in connected status (Standby mode).
Image creation is possible.
2) Is the data acquisition mode and accumulation time etc. usage is set?
Please verify the use setting and set.
Question
Accumulation time setting is not effective.
Answer
Is the data acquisition mode set in external trigger level mode
("EXT.LEVEL1"mode) ?
If the data acquisition mode in use is other than the external trigger level
mode, then set the data acquisition mode to be used (internal synchronous
mode ("INT" mode), external trigger edge mode ("EXT.EDGE1" mode)).
* If the data acquisition mode is in use is external trigger level mode, then
accumulation time setting is not required and so there will not be a problem
if it is not effective.
Question
Data acquisition mode cannot be set as external trigger edge mode.
Answer
Is the accumulation time set less than 40 msec?
Set the accumulation time more than 41 msec.
Question
Accumulation time cannot be set below 40 msec.
Answer
Is the data acquisition mode set in external trigger edge mode?
In external trigger edge mode, the minimum accumulation time is 41 msec.
It is not possible to set below 40 msec.