2
Product information
Product name
Thingsee AIR
Product description
Indoor AIR quality measurement
Vendor name
Haltian Products Oy
Product type
Wirepas sensor
Internal Model Code (IMC prefix for TUID)
•
TSAR01
•
XXAR01
Sensor measurement specifications
CO₂
•
Measure CO₂ levels indoors
•
Measurement range 400 ... 5000 ppm
•
Accuracy is ±(30ppm +3% of reading)
•
Automatic baseline calibration takes 8 days, and requires that the CO₂
level drops at least once per day
•
Possibility for manual baseline calibration
•
CO₂ sensor is a nondispersive infrared sensor (NDIR)
TVOC
•
Total volatile organic compounds
•
Resolution is 1ppb at range 0ppb-2008ppb
•
Resolution is 6ppb at range 2008ppb – 11110ppb
•
Resolution is 32 ppb at range 11110ppb - 60000ppb
•
Automatic baseline calibration takes 24 hours.
Temperature
•
Measurement range -40 ... 85 °C
•
Accuracy is ±0.2°C
Relative humidity
•
Accuracy is ±2%
Barometric
pressure
•
Measurement range 300 ... 1200 hPa
•
Absolute accuracy ±1hPa
•
Barometric sensor precision ±0.002hPa
Ambient light
•
Measurement is not calibrated
•
Use only for indicative purposes