Instruction Manual REG 21
21
7
Troubleshooting
The instrument recognizes the following problems and responds by returning an error
message:
Error
message
Problem
Cause
Corrective Action
E003
overloaded
pressure
measurement
capsule
pressure too high
send the instrument to the
manufacturer for repair
E004
pressure
measurement
capsule is
pressurized during
zero-point
calibration
pressure has not been
disconnected
defective pressure
measurement capsule
reduce pressure
send the instrument to the
manufacturer for repair
Table 6