Instruction Manual PUC 28
27
Now a listing of all commands for configuration:
No.
Name
Format
Unit
Description
1
ScalPdown
float
Pa
Scaling of int. pressure, bottom value (for
analogue output)
2
ScalPup
float
Pa
Scaling of int. pressure, top value (for analogue
output)
3
ScalIN1Tdown
float
°C
Scaling of IN1, bottom value (input IN1 as
temperature)
4
ScalIN1Tup
float
°C
Scaling of IN1, top value (input IN1 as
temperature)
5
ScalIN1Pdown
float
Pa
Scaling of IN1, bottom value (input IN1 as
pressure)
6
ScalIN1Pup
float
Pa
Scaling of IN1, top value (input IN1 as pressure)
7
ScalIN2Hdown
float
%rH
Scaling of IN2, bottom value (input IN2 as
humidity)
8
ScalIN2Hup
float
%rH
Scaling of IN2, top value (input IN2 as humidity)
9
ScalIN2Pdown
float
Pa
Scaling of IN2, bottom value (input IN2 as
pressure)
10
ScalIN2Pup
float
Pa
Scaling of IN2, top value (input IN2 as pressure)
11
WarnPdown
float
Pa
Warning of int. pressure, bottom value
12
WarnPup
float
Pa
Warning of int. pressure, top value
13
WarnPHyst
float
Pa
Warning of int. pressure, hysteresis
14
WarnIN1Tdown
float
°C
Warning of IN1, bottom value (input IN1 as
temperature)
15
WarnIN1Tup
float
°C
Warning of IN1, top value (input IN1 as
temperature)
16
WarnIN1THyst
float
°C
Warning of IN1, hysteresis (input IN1 as
temperature)
17
WarnIN1Pdown
float
Pa
Warning of IN1, bottom value (input IN1 as
pressure)
18
WarnIN1Pup
float
Pa
Warning of IN1, top value (input IN1 as
pressure)
19
WarnIN1PHyst
float
Pa
Warning of IN1, hysteresis (input IN1 as
pressure)
20
WarnIN2Hdown
float
%rH
Warning of IN2, bottom value (input IN2 as
humidity)
21
WarnIN2Hup
float
%rH
Warning of IN2, top value (input IN2 as
humidity)
22
WarnIN2HHyst
float
%rH
Warning of IN2, hysteresis (input IN2 as
humidity)
23
WarnIN2Pdown
float
Pa
Warning of IN2, bottom value (input IN2 as
pressure)
24
WarnIN2Pup
float
Pa
Warning of IN2, top value (input IN2 as
pressure)
In
case the PUC 28 doesn’t accept the value which has been sent, the
answer contains an error code (see section 6.2).
A value which has been sent is effective immediately, but it is not saved
permanently automatically. For this, the command
“SaveSet” has to be used.