halstrup-walcher PS27 Operating Instructions Manual Download Page 8

 

Operating Instructions PS27 

 

3.3 Connections for Two-Wire System (Option): 

 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

One measurement range  

 

 

Four measurement ranges 

 
 
 

ST4 Pin 

Connection name 

Meaning 

24V 

+terminal 15-32VDC 

GND 

Not used 

OUT 

-terminal 

GND 

Not used 

SET 0 

Zero-point input 

 

4. 

Resetting Offset

 

 

After starting the pressure transducer, give it some 15 minutes to warm 
up. During this time the output signal may be unstable. 

 

Zero-point calibration of the appliance is only effected via the input assigned 

for this purpose. 

The user must ensure that the appliance is not subjected to any 

pressure during the calibration process as this might lead to a wrong offset 
value which, in turn, would result in a wrong output signal. 

The 24V input makes 

it possible to activate a 3/2 directional control valve with the same signal, which 
switches the PS27's pressure inputs at zero pressure during the zero calibration 
process. In this case the zero point calibration signal must be applied until there is no 
doubt that the pressure has fallen to zero.  
 

 

While Resetting Offset is in progress, the power supply must not be 

switched off. This might result in a loss of the device setting and may 

leave the PS27 unusable. 

 

 

5  4  3  2  1  

5  4  3  2  1  

Summary of Contents for PS27

Page 1: ...ial Pressure Transducer PS27 halstrup walcher GmbH Stegener Stra e 10 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E mail info halstrup walcher com Internet www halstrup...

Page 2: ...Transducer 6 3 Startup 6 3 1 Functional Description 6 3 2 Connections for Four Wire version standard in the device 7 3 3 Connections for Two Wire System Option 8 4 Resetting Offset 8 5 Configuring th...

Page 3: ...service life of the transducer They must be passed on to the succeeding personnel charged with operations Any additions coming from the manufacturer must be inserted The manufacturer reserves the rig...

Page 4: ...sons as assigned by the plant operator are to operate the transducer Do not perform any functional test involving compressed or breathable air This would otherwise damage those transducers with low me...

Page 5: ...ndling the equipment WARNING It points to a hazard possibly extending from injures through to death should the instructions not be followed CAUTION It points to a hazard which could cause considerable...

Page 6: ...as this may lead to measurement errors arising Ideally the appliance should be vertically mounted on a wall not subject to any vibrations To stop any condensed water entering the measurement cell the...

Page 7: ...gram on the appliance cover circuit board mount The transducer outputs are protected from short circuits The DC supplied appliances are protected from reverse polarity of the supply voltage ST4 Pin Co...

Page 8: ...ser must ensure that the appliance is not subjected to any pressure during the calibration process as this might lead to a wrong offset value which in turn would result in a wrong output signal The 24...

Page 9: ...aching of 4 measuring ranges into the appliance has been undertaken by the manufacturer The switches 1 and 2 of SW can be used to select the measuring range in question before the transducer is switch...

Page 10: ...t so that any adjusted time delay will also cause a delay in relay triggering 9 In case of Malfunctions Fault description Possible cause Corrective action No output signal Supply voltage is not connec...

Page 11: ...500 Ohm 4 20 mA two wire 17 32 V RL 500 Ohm Switching contact optional max 230 VAC 6 A min 500 mW Display optional 3 digit LCD Overload range Range 2 0kPa 10 kPa Range 10 kPa 200kPa Range 100 kPa 500k...

Page 12: ...27 12 11 Dimensional Drawing For additional specifications and dimension drawings please visit our website at http www halstrup walcher de en products measurement technology index php 7100 004354I_PS2...

Page 13: ...Operating Instructions PS27 13 12 Certificate of Conformity...

Reviews: