Troubleshooting 

Error description 

Potential cause 

Corrective action 

no output signal 

• supply voltage is not connected 
• incorrect supply voltage 
• defective reverse polarity 
protection diode 
• defective input protection diode 

• connect correct supply voltage 
• connect the correct supply 
voltage (see rating plate). 
• send the instrument to the 
manufacturer for repair 
• send the instrument to the 
manufacturer for repair 

output signal is constant, 
despite change in 
pressure 

• defective output protection diode 
• pressure ports reversed 

• send the instrument to the 
manufacturer for repair 
• connect the pressure as 
described in Section 3.1 

output signal incorrect 

• defective output protection diode 
• defective pressure measurement 
cell 

• send the instrument to the 
manufacturer for repair 
• send the instrument to the 
manufacturer for repair 

zero point cannot be 
adjusted using P1 

• defective pressure measurement 
cell 

• send the instrument to the 
manufacturer for repair 

 

7  

Technical data 

Measurement data 

 

measurement ranges 

±300 Pa to 200 kPa 

overload capacity 

10 x the final value of the measurement range 
(for measurement ranges 

 20 kPa) 

20 x the final value of the measurement range 
(for measurement ranges > 20 kPa) 

hysteresis 

0.1 % 

warm-up period 

approx. 30 min. 

time required for adjustment 

approx. 20 ms (up to 5 s available upon request) 

deviation from characteristic curve 
(setting limiting value) 

2 % 

temperature-dependent drift in zero 
point 

0.1%/ K (within the +10°C...+50 °C range) 

temperature-dependent drift in 
measurement range 

0.1%/ K (within the + 10°C...+50 °C range) 

dead volume 

approx. 2000 mm

3

 (for measurement ranges 

 250 Pa) 

control volume 

max. 200mm³ 

max. system pressure 

100 kPa 

Ambient conditions 

 

medium 

air, all non-aggressive gases 

nominal temperature 

+10° C to +50° C 

operating temperature 

0° C to +70° C 

storage temperature 

-10° C to +70° C 

relative humidity 

0…80 % 

EMC standards 

corresponds to EN 50 081 part 1 and EN 50 082 part 1 

conformity 

 
declaration of conformity available upon request 

Summary of Contents for PS11

Page 1: ...l Mounting Type PS11 Type PK11 with Contact Point halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E mail info halstrup walcher com Interne...

Page 2: ...ctions and start up 4 1 3 Troubleshooting maintenance repairs disposal 4 1 4 Symbols 5 2 Instrument description 6 3 Start up 6 3 1 Features 6 3 2 Supply voltage connections in the instrument 7 4 Calib...

Page 3: ...ly available throughout the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary...

Page 4: ...ised by the operator of the facility The instrument may only be operated by appropriately trained individuals who have been authorized by the operator of the facility Do not carry out a function test...

Page 5: ...r operation could result in the following hazards WARNING This warns you of a potential hazard that could lead to bodily injury up to and including death if the corresponding instructions are not foll...

Page 6: ...ictional parts or parts subject to mechanical wear 3 Start up 3 1 Features Although the PS11 PK11 pressure transducer is highly robust it is nevertheless a precision instrument and should be handled w...

Page 7: ...uments are calibrated by the manufacturer before delivery During assembly it should only be necessary to correct any positional errors However the following tools are required if calibration is still...

Page 8: ...point with the NO contact function The same instructions apply to setting the NC contact function only the function is inverted You can access the P6 potentiometer through a bore in the front of the i...

Page 9: ...defective pressure measurement cell send the instrument to the manufacturer for repair 7 Technical data Measurement data measurement ranges 300 Pa to 200 kPa overload capacity 10 x the final value of...

Page 10: ...an be adjusted over the complete measurement range using the potentiometer Physical data pressure port FESTO elbow AN quick connector for PVC tubing external 6mm electrical connection screw terminals...

Page 11: ...ion drawings 111 45 U OUT 20 I OUT 19 5 UB GND GND 31 25 73 111 Verschlu stopfen f r Schlauch au en 6 2x L Steckverschraubung 7100 003754B_PS11_PK11_eng doc 02 11 2021 Ze cover seal FESTO elbow AN qui...

Reviews: