Instruction manual PS11/PK11
9
6
Troubleshooting
Error description
Potential cause
Corrective action
no output signal
• supply voltage is not connected
• incorrect supply voltage
• defective reverse polarity
protection diode
• defective input protection diode
• connect correct supply voltage
• connect the correct supply
voltage (see rating plate).
• send the instrument to the
manufacturer for repair
• send the instrument to the
manufacturer for repair
output signal is constant,
despite change in
pressure
• defective output protection diode
• pressure ports reversed
• send the instrument to the
manufacturer for repair
• connect the pressure as
described in Section 3.1
output signal incorrect
• defective output protection diode
• defective pressure measurement
cell
• send the instrument to the
manufacturer for repair
• send the instrument to the
manufacturer for repair
zero point cannot be
adjusted using P1
• defective pressure measurement
cell
• send the instrument to the
manufacturer for repair
7
Technical data
Measurement data
measurement ranges
±300 Pa to 200 kPa
overload capacity
10 x the final value of the measurement range
(for measurement ranges
20 kPa)
20 x the final value of the measurement range
(for measurement ranges > 20 kPa)
hysteresis
0.1 %
warm-up period
approx. 30 min.
time required for adjustment
approx. 20 ms (up to 5 s available upon request)
deviation from characteristic curve
(setting limiting value)
2 %
temperature-dependent drift in zero
point
0.1%/ K (within the +10°C...+50 °C range)
temperature-dependent drift in
measurement range
0.1%/ K (within the + 10°C...+50 °C range)
dead volume
approx. 2000 mm
3
(for measurement ranges
250 Pa)
control volume
max. 200mm³
max. system pressure
100 kPa
Ambient conditions
medium
air, all non-aggressive gases
nominal temperature
+10° C to +50° C
operating temperature
0° C to +70° C
storage temperature
-10° C to +70° C
relative humidity
0…80 %
EMC standards
corresponds to EN 50 081 part 1 and EN 50 082 part 1
conformity
declaration of conformity available upon request