background image

 

PU/PI and PS10 Instruction Manual 

11 

 

Electrical data

 

 

power consumption 

max. 1.2 W 

supply voltage 

24 VDC +20% / -15% 
(filtered, permissible peak-to-valley ratio = 1000 mV) 
 
230VAC, 115VAC, 24VAC +6 % / -15 %, 50/60 Hz 
(optional) 

minimum load resistance 

R

L

 

R

L

  

 2 k

 for an output voltage of 0...10 V 

R

L

  

 5 k

 for an output voltage of 

5 V 

R

L

  

 10 k

 for an output voltage of 

10 V 

for models PU and PS10 with output voltage 
 
maximum effect of load resistance = 0.3% 

maximum output load 

R

B

 

R

B

 

=

 500 

 

for models PI and PS 10 with current output 
 
output load dependence: < 0.3 % 

display 

3½ or 4½-place LCD, character height = 13 mm 
(optional) 

output signal 

0 to 10 V, 0 to 20 mA or 4 to 20 mA; ±5 V and ±10 V are 
also possible 

 

 

Physical data

 

 

pressure port 

Ø 6.5 mm for NW5 tubing 
(interior tubing diameter = 5 mm) 

electrical connections 

screw terminals for cables up to 2.5 mm

2

 

mounting orientation 

vertical 
(when placing your order, please indicate if a horizontal 
orientation is required) 

dimensions (w x h x d) 

80 x 120 x 73 mm 
120 x 122 x 75 mm (for LCD option) 
 
 

protection class 

IP54 

weight 

between 300 g and 700 g, depending on design 

options 

 3½-place LCD 

 4½-place LCD 

 linearity protocol 

 DKD calibration certificate 

 cable gland: PG9 or PG11 threaded connections 

 output signal attenuation up to 5 s 

 silicon-free materials for parts in contact with medium 

Appendix A: Parts in contact with measurement medium 

 Beryllium bronze  CuBe2 

 Araldite CY236 / HY988 

 Mu metal (nickel alloy) 

 Loctite 242e 

 Brass  CuZn39Pb3 

 Carbonyl iron 

 Aluminum  AlCuMgPb / AlMg3 

 KEL  (FPM: fluorinated rubber)  

 Silicon (tubing) optional: Viton 

 Vepuran Vu 4457/51 

 Crastin (PTBP) 

 UHU-Plus endfest 300 binder 

Summary of Contents for PI10

Page 1: ...Pressure Transducers halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E Mail info halstrup walcher com Internet www halstrup walche...

Page 2: ...ctions and start up 4 1 3 Troubleshooting maintenance repairs disposal 4 1 4 Symbols 5 2 Instrument description 6 3 Start up 6 3 1 Features 6 3 2 Supply voltage connections in the instrument 7 3 3 Ana...

Page 3: ...e throughout the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials p...

Page 4: ...ely trained and authorized by the operator of the facility may assemble the instrument and set up its electrical connections The instrument may only be operated by appropriately trained individuals wh...

Page 5: ...The symbols given below are used throughout this manual to indicate instances when improper operation could result in the following hazards WARNING This warns you of a potential hazard that could lead...

Page 6: ...theless precision instruments and should be handled with care Avoid mounting the instruments in the direct vicinity of any sources of radiation or heat such as heaters as this could result in measurem...

Page 7: ...connection diagram located on the housing cover circuit board mount 3 3 Analog output connections in the instrument The transducer outputs are protected from short circuits Instruments supplied with d...

Page 8: ...ark when calibrating the zero point Were this not the case the negative component which cannot be displayed would interfere with the output signal After the pressure transducer has warmed up the opera...

Page 9: ...de pressure ports reversed Send the instrument to the manufacturer for repair connect pressure as outlined in section 3 Mounting output signal incorrect only for 0 20 mA defective output protection di...

Page 10: ...s 250 Pa optional 2 for PS 10 model others available upon request temperature dependent drift in zero point 0 04 K in the 10 C 50 C range for PU PI model 0 1 K in the 10 C 50 C range for PS 10 model t...

Page 11: ...6 5 mm for NW5 tubing interior tubing diameter 5 mm electrical connections screw terminals for cables up to 2 5 mm2 mounting orientation vertical when placing your order please indicate if a horizont...

Page 12: ...PU PI and PS10 Instruction Manual 12 7 Dimension drawings PU PI PS10 standard 120 135 151 20 80 25 33 7 4 5 34 73 PU PI PS10 with LCD 122 17 25 22 75 Option LCD 4 5 111 90 5 20 120...