background image

 

P26 Instruction Manual 

13 

 

4 Zero-point calibration cycle 

External influences such as temperature, position or ambient pressure can shift the instrument’s 

zero point, i.e., the value displayed when the pressure ports are open. Calibration is the process 

by which the instrument automatically registers this shift and figures it into the currently 

displayed pressure value. The zero-point calibration is performed in two stages, which are 

shown in the top line on the display (optional). 

 0” Measuring signal of the zero point is being determined. 

 P” Pressure is being reconfigured 

The interval between two zero-point calibrations can be adjusted in the Settings menu.  

Zero-point calibration can also be started with the Enter key if the device is in Measurement 

mode. 

The P26 does not respond to keys being pressed during calibration. 

4.1 External zero-point calibration(Option) 

The ext. zero-point calibration is available as an Option.  To initiate an ext. zero-point calbration, 

the extra pin on the output connector must be connected to +24 V.  

 

INFORMATION: 
 

If a zero-point calibration is initiated via the serial interface or optional ext. zero-point 

calibration, this will still be carried out. In this case, it doesn’t matter if the zero-point 

calibration is deactivated or the P26 is in the menu. 

 

 

5 Overpressure protection 

The P26 has an internal overpressure safeguard that protects the precision pressure 

measurement capsule from damage. (Overload range: 200 x [max. 600 kPa]) 

6 Display (optional) 

The measured values and the menu are shown on the display. The top line is the information 

line. This line displays the type of measured value or the status of a zero-point calibration. The 

measurement value will be displayed on the central line. If the permitted measuring range is 

exceeded or not reached, the figures are replaced by corresponding arrows. The bottom line 

displays the unit, with the first two digits showing the status of the two relays.  

7 Menu (optional) 

If the P26 includes a display, you can use the menu to adjust various settings. Press the Menu 

key to bring up the menu. 

In Menu mode, the parent menu item always appears in the top display line. The middle line 

always shows the current sub-menu item or value to be changed. The bottom line displays units 

or other help texts.  

Start the Menu mode by pressing the "Menu" key. It can be protected with a four-character 

password. At the main menu level, “Menu” is displayed in the top line and “Display” in the 

middle line. Select your desired sub-menu by pressing the “^” and “>” keys. Press “Enter” to go 

to a sub-menu or enter a value. Press the “Menu” key to go to the next menu level or to stop 

entering values.  

 

 

Summary of Contents for P26

Page 1: ...Pressure Transducer halstrup walcher GmbH Stegener Strasse 10 D 79199 Kirchzarten GERMANY Phone 49 7661 39 630 Fax 49 7661 396 399 E Mail info halstrup walcher com Internet www halstrup walcher com Do...

Page 2: ...nal ports and keys 9 3 4 Front view 12 4 Zero point calibration cycle 13 5 Overpressure protection 13 6 Display optional 13 7 Menu optional 13 7 1 Display 14 7 2 Scale 14 7 2 1 Pressure 14 7 2 1 1 Top...

Page 3: ...17 7 4 4 Warning signal 17 7 4 5 Resolution 18 7 4 6 Zero point calibration 18 7 4 7 CS creep suppression 18 7 4 8 Read factory settings 18 7 4 9 Code 18 8 Serial interface optional 18 8 1 Settings 18...

Page 4: ...the lifecycle of the instrument It must be provided to any individuals who assume responsibility for operating the instrument at a later date It must include any supplementary materials provided by t...

Page 5: ...trained individuals who have been authorized by the operator of the facility Do not carry out a function test with compressed or breathable air This would damage instruments with low measuring ranges...

Page 6: ...rns you of a potential hazard that could lead to bodily injury up to and including death if the corresponding instructions are not followed CAUTION This warns you of a potential hazard that could lead...

Page 7: ...Measurement of differential pressure Measurement of volumetric flow mass flow and flow rate Display of a measured value Monitoring of a variety of threshold parameters Actual pressure value Calculati...

Page 8: ...ation Confirm Up Max display Increase value or scroll up Right Min display Move the cursor or scroll down In measurement mode the keys perform the following functions The Enter key starts a zero point...

Page 9: ...P26 Instruction Manual 9 3 3 Internal ports and keys Standart version with external zero point calibration...

Page 10: ...l 10 Standard Version Output signal Relay 1 Relay 2 Power supply RS232 interface for PC Reset Version mit externer Nullierung Output signal Relay 1 Relay 2 Power supply ext offset calibration RS232 in...

Page 11: ...interface for PC RS232 or USB Interface interface for parameter settings Reset The processor is restarted Port description Description 1 OUT_I Current output 2 OUT_GND Earth 3 OUT_U Voltage output Z...

Page 12: ...3 Reception TXD 2 Send GND 5 Earth USB Port description PIN Description Power 1 Power supply for USB from PC D bidirectional 2 USB Data negative polarity D bidirectional 3 USB Data positive polarity...

Page 13: ...e zero point calibration is deactivated or the P26 is in the menu 5 Overpressure protection The P26 has an internal overpressure safeguard that protects the precision pressure measurement capsule from...

Page 14: ...ndicates the output value This pressure range forms the basis for many other settings such as the limits for the relays warnings and factors for the flow display There are two default settings for the...

Page 15: ...s The following sub menu items are available Value Unit 7 2 3 1 Value This is used to set the mass flow value displayed at the maximum scaled pressure 0 is always used as the lower scale value on the...

Page 16: ...y is displayed select the Pa unit 7 3 2 Hysteresis This item can be used to select the hysteresis for the particular relay The action described in 7 3 1 applies to the entry unit The hysteresis is alw...

Page 17: ...put Filter Warning signal Resolution Zero point calibration CS creep suppression Read factory settings Code 7 4 1 Language The menu language can be selected with this menu item You can choose from the...

Page 18: ...works with volumetric flow mass flow and flow rate root extracted measurement values 7 4 8 Read factory settings The factory settings are established in the instrument before delivery With this menu i...

Page 19: ...nit 120 120 from measurement range floating RRelai1 Switching direction for Relay 1 1 decreasing 0 off 1 increasing int SRelai1 Sound for Relay 1 ms 0 Off max 1000ms HRelai1 Hysteresis relay 1 Pa 0 12...

Page 20: ...int DAC Out Analogous output signal 0 4 20mA 1 0 20mA 2 5 to 5V 3 0 10V unsigned int Res Resolution 0 max 1 0 01 2 0 1 3 0 2 4 0 5 5 1 unsigned int UnitD Unit display 0 Pa 1 hPa 2 kPa 3 mbar 4 mmH2O...

Page 21: ...ltage See rating plate Output signal 0 10 V RL 2 k or 5 5 V RL 2 k or 0 20 mA RL 500 or 4 20 mA RL 500 Relays 2 x change over contacts For each relay 6A 230 VAC Interface RS 232 9600 Baud 8 data bits...

Page 22: ...P26 Instruction Manual 22 11 Dimension drawing 147 5 1 5 56 5 1 5 25 39 5 20 92 5 162 100 5 6 5 6 5 4 5 14 5 27 75...

Page 23: ...mH2O mmHg psi inH2O inHg m s m h kg s kg min kg h m s mph f s f min km h Scaling Pressure Top xxxx x Bottom xxxx x Unit Pa hPa kPa mbar mmH2O mmHg psi inH2O inHg Vol Flow Value xxxxxxx x Unit m s m h...

Page 24: ...11 xxxxx ms Warning signal xxxxx ms Filter Rev2 14 off on Mode increasing decreasing Settings Language English German Italian French Output 4 20 mA 0 20 mA 5 5V 0 10 V Filter xxxxx ms Warning signal...

Reviews: